Midbrook Inc.
2080 Brooklyn Rd, Jackson, Michigan, United States
FEI 3000232301
FDA has 2 inspections on file for Midbrook Inc. in Jackson, Michigan (FEI 3000232301), from 2012 to 2014. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 16 citations from this inspection: 15 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 806, the device correction and removal reporting rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 2, 2012–2014
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DMR - not or inadequately maintained
A device master record has not been maintained.
Lack of or inadequate instructions
test procedures have not been adequately established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Lack of procedures, or not maintained
Document control procedures have not been established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals and with sufficient frequency.
Show 11 more citations
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Records not kept
There is no record maintained of a correction or removal action that was not required to be reported to FDA.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Change records, content
Records of changes did not include a description of the change, identification of the affected documents, and the signature of the approving official(s).
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.
Design validation - documentation
The results of design validation, including identification of the design, method(s), the date, and the individual(s) performing validation, were not documented in the design history file.
Design validation - software validation not performed
Validation of device software is incomplete.
Design review - documentation
The design review results, including identification of the design, the date, and the individual(s) performing the review, were not documented in the design history file.
Show 8 more citations
Product rework adverse effects {see also 820.184}
Documentation of rework and reevaluation activities does not include a determination of whether there has been any adverse effect from rework upon the product.
DHR documentation of label release {see also 820.184}
The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been established.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Complaints
Complaint files are not adequately maintained.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.25Personnel2 of 2
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.40[Reserved]2 of 2
- 21 CFR 820.90Nonconforming product2 of 2
- 21 CFR 820.181Device master record2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 2
- 21 CFR 820.20Management responsibility1 of 2
Show all 16
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.120Device labeling1 of 2
- 21 CFR 820.170Installation1 of 2
- 21 CFR 820.184Device history record1 of 2
- 21 CFR 820.200Servicing1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3000232301 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3000232301. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.