Pharmaceutical Innovations, Inc.
897 Frelinghuysen Ave, Newark, New Jersey, United States
FEI 2241583
FDA has 5 inspections on file for Pharmaceutical Innovations, Inc. in Newark, New Jersey (FEI 2241583), from 2011 to 2019. The latest ended on and was classified OAI, official action indicated.
OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.
FDA published 11 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
FDA issued a warning letter on (CDRH). FDA's compliance record also lists 2 other actions under this FEI, the latest Injunction, . 2 import refusals are filed under this FEI, the latest on .
Seven regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- OAI
- Inspections on file
- 5, 2011–2019
- 5 OAI · 0 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- 2 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate In-process acceptance procedures
Procedures for the acceptance and control of in-process product have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Design history file
The design history file was not established.
Implementing Personnel Procedures, Health, Cleanliness.
Requirements addressing the personal practices of employees have not been implemented.
Show 6 more citations
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Lack of procedures, or not maintained
Document control procedures have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been established.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers and contractors have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a specification and process have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation - production units
The design was not validated under defined operating conditions and using initial production units, lots or batches or their equivalents.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Distribution records
Distribution records were not maintained and do not include or refer to the location of required information.
Design validation - production units
The design was not validated using initial production units, lots or batches or their equivalents.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a specification and procedure have not been adequately established.
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
Show 6 more citations
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Availability
Required records were not made readily available for review and copying by the FDA and are not legible.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Documented evaluation
The evaluation of potential contractors was not documented.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.70Production and process controls4 of 4
- 21 CFR 820.30Design controls3 of 4
- 21 CFR 820.50Purchasing controls3 of 4
- 21 CFR 820.75Process validation3 of 4
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 4
- 21 CFR 820.90Nonconforming product2 of 4
- 21 CFR 820.181Device master record2 of 4
Show all 14
- 21 CFR 820.184Device history record2 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
- 21 CFR 820.160Distribution1 of 4
- 21 CFR 820.180General requirements1 of 4
- 21 CFR 820.198Complaint files1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- InjunctionCDRH · DevicesFDA record
- SeizureCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2241583, one row per refused product line.
2 refusals from 2018-05-30 to 2018-09-21 · charged STERILITY on 2 · FDA sample analysis on 1
FDA firm profile- 2018-09-21
MEDIA, COUPLING, ULTRASOUND90MUI
STERILITY 501(a)(2)(A), 801(a)(3); ADULTERATION
Division of Southwest ImportsBJV-4094321-0/1741/1
- 2018-05-30
MEDIA, COUPLING, ULTRASOUND90MUI
STERILITY 501(a)(2)(A), 801(a)(3); ADULTERATION
FDA sample analysis
Division of Southwest ImportsBST-0186968-5/181/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2241583 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2241583. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.