Skip to content

Pharmaceutical Innovations, Inc.

897 Frelinghuysen Ave, Newark, New Jersey, United States

FEI 2241583

FDA has 5 inspections on file for Pharmaceutical Innovations, Inc. in Newark, New Jersey (FEI 2241583), from 2011 to 2019. The latest ended on and was classified OAI, official action indicated.

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 11 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

FDA issued a warning letter on (CDRH). FDA's compliance record also lists 2 other actions under this FEI, the latest Injunction, . 2 import refusals are filed under this FEI, the latest on .

Seven regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
5, 2011–2019
5 OAI · 0 VAI · 0 NAI
Inspections with 483 observations shown
4 of 5
Form 483s posted by FDA
None held for this FEI
Import refusals
2 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate In-process acceptance procedures

      Procedures for the acceptance and control of in-process product have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    4. Design history file

      The design history file was not established.

    5. Implementing Personnel Procedures, Health, Cleanliness.

      Requirements addressing the personal practices of employees have not been implemented.

    Show 6 more citations
    1. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    2. Lack of procedures, or not maintained

      Document control procedures have not been established.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    4. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    6. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Contamination control, Lack of or inadequate procedures

      Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been established.

    5. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers and contractors have not been adequately established.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a specification and process have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Design validation - production units

      The design was not validated under defined operating conditions and using initial production units, lots or batches or their equivalents.

    4. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    2. Distribution records

      Distribution records were not maintained and do not include or refer to the location of required information.

    3. Design validation - production units

      The design was not validated using initial production units, lots or batches or their equivalents.

    4. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a specification and procedure have not been adequately established.

    5. Evaluation, timeliness, identification

      Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.

    Show 6 more citations
    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    2. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    3. Availability

      Required records were not made readily available for review and copying by the FDA and are not legible.

    4. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    5. Documented evaluation

      The evaluation of potential contractors was not documented.

    6. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.

Show all 14

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • InjunctionCDRH · DevicesFDA record
  • SeizureCDRH · DevicesFDA record
  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2241583, one row per refused product line.

2 refusals from 2018-05-30 to 2018-09-21 · charged STERILITY on 2 · FDA sample analysis on 1

FDA firm profile
  • 2018-09-21

    MEDIA, COUPLING, ULTRASOUND90MUI

    STERILITY 501(a)(2)(A), 801(a)(3); ADULTERATION

    Division of Southwest ImportsBJV-4094321-0/1741/1

  • 2018-05-30

    MEDIA, COUPLING, ULTRASOUND90MUI

    STERILITY 501(a)(2)(A), 801(a)(3); ADULTERATION

    FDA sample analysis

    Division of Southwest ImportsBST-0186968-5/181/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2241583 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2241583. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.