Danmar Products, Inc.
221 Jackson Industrial Dr, Ann Arbor, Michigan, United States
FEI 1831680
FDA has 6 inspections on file for Danmar Products, Inc. in Ann Arbor, Michigan (FEI 1831680), from 2009 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 7 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Jan 15, 2009. No warning letter, court action or import refusal is filed under this FEI.
Eight regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2009–2024
- 1 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 6 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Apr 30, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documented evaluation
The evaluation of potential suppliers was not documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Documentation
Process validation results have not been adequately documented.
Training records
Personnel training is not documented.
Show 2 more citations
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, checked and maintained have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Documentation
Acceptance activities were not adequately documented.
Records not kept
There is no record maintained of a correction or removal action that was not required to be reported to FDA.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 2 more citations
Documentation
Corrective and preventive action activities and/or results have not been documented.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Document review, approval by designated individual
Documents were not approved by designated individual(s) prior to issuance .
Training records
Personnel training is not documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been established.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been established.
Personnel
Personnel do not have the necessary training to perform their jobs.
Show 1 more citation
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design history file
The design history file was not established or maintained.
ID label, labeling
The device history record does not include the primary identification label and labeling for each device.
Acceptable supplier records
Records of acceptable suppliers were not maintained.
CAPA verification/validation of corrective/preventive action
Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.
Design transfer - Lack of or inadequate procedures
Procedures were not defined to ensure that the device design was correctly transferred into production.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Incoming acceptance records, documentation
Acceptance or rejection of incoming product was not documented.
Procedures
The procedures addressing documentation of corrective and preventive action activities were not complete.
Management representative
No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
Management review - Lack of or inadequate procedures
Procedures for management review are not established.
Documentation
Acceptance test results of in-process product were not fully documented.
Show 8 more citations
Uniform and timely processing
Complaint handling procedures have not been completed and implemented to ensure that all complaints are processed in a uniform and timely manner.
Documentation
Acceptance activities were not fully documented.
Design history file
The design history file was not established or maintained.
Quality audits - Lack of or inadequate procedures
Procedures for conducting quality audits were not established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed and implemented.
Lack of or inadequate procedures
The procedures for implementing corrective and preventive actions were not complete and implemented.
DMR QA procedures and specifications
The device master record does not include or refer to the location of quality assurance procedures and specifications.
Supplier notification of changes
There is no clear agreement from the suppliers that they will notify you of changes in the product or service.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Apr 30, 2025. It asked for a Form 483.
Inspection ending Nov 4, 2024VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Apr 30, 2025FDA-FOIA-2025-3887Form 483· 1 of 4 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action5 of 6
- 21 CFR 820.50Purchasing controls4 of 6
- 21 CFR 820.25Personnel3 of 6
- 21 CFR 820.30Design controls3 of 6
- 21 CFR 820.72Inspection, measuring, and test equipment3 of 6
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 of 6
- 21 CFR 820.22Quality audit2 of 6
- 21 CFR 820.198Complaint files2 of 6
Show all 16
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 6
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 6
- 21 CFR 820.20Management responsibility1 of 6
- 21 CFR 820.40[Reserved]1 of 6
- 21 CFR 820.75Process validation1 of 6
- 21 CFR 820.90Nonconforming product1 of 6
- 21 CFR 820.181Device master record1 of 6
- 21 CFR 820.184Device history record1 of 6
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1831680 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1831680. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.