The Kilroy Company
9470 Pinecone Dr, Mentor, Ohio, United States
FEI 1527214
FDA has 3 inspections on file for The Kilroy Company in Mentor, Ohio (FEI 1527214), from 2009 to 2011. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
11 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 3, 2009–2011
- 0 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices17 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.
Lack of or inadequate DHR procedures
Procedures have not been defined and implemented to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
DMR device specifications
The device master record does not include or refer to the location of all specifications.
CAPA verification/validation of corrective/preventive action
Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.
Document review, approval by designated individual
Documents were not reviewed and approved by the individual designated in document control procedures.
Show 12 more citations
Procedures for product review,disposition lack of/inadequate
Procedures that describe the review and disposition process for nonconforming product were not implemented.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy has been fully implemented and maintained at all levels of the organization.
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been established.
Documentation of disposition, justification, signature
There is incomplete documentation of the disposition of nonconforming product, justification for use of nonconforming product, and the signature of the individual authorizing the use of nonconforming product.
Equipment
Acceptance records did not identify the equipment used in the tests.
Procedure for analysis of data sources
The corrective and preventive procedures addressing the analysis of sources of quality data to identify existing and potential causes of nonconforming product or other quality problems were not implemented.
Appropriateness
There is no information to support the appropriateness of the statistical techniques used.
Uniform and timely processing
Complaint handling procedures have not been established and defined to ensure that all complaints are processed in a uniform and timely manner.
Sampling plans
Sampling plans are not based on valid statistical rationale.
Documentation
Acceptance activities were not fully documented.
QSR
The quality system record does not include or refer to the location of quality system procedures and documentation of quality system activities.
Software validation for automated processes
Software used as part of production has not been validated for its intended use according to an established protocol.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices35 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Management review - defined interval, sufficient frequency
Management reviews were not conducted at defined intervals.
Not approved or obsolete document retrieval
Documents that were obsolete were observed at a location where they could be used.
Sampling plans
Sampling plans are not defined, documented, completed, implemented, and based on valid statistical rationale.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
Equipment suitability & capability
Certain inspection equipment is not suitable for its intended purposes or capable of producing valid results.
Show 30 more citations
Documentation
Acceptance activities were not fully documented.
Document review, approval by designated individual
Documents were not reviewed and approved by the individual designated in document control procedures.
Specific non-conforming product procedures
Procedures for addressing the investigation of nonconforming product were not defined and documented.
Lack of procedures, or not maintained
Document control procedures were not complete and implemented.
Acceptance records
The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.
Quality policy and objectives
The quality policy and objectives were not complete.
Management responsibility
Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.
Investigation
Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to specifications, methods, processes, and procedures were not documented, complete, and implemented.
Supplier oversight
The type and extent of control to be exercised over services and suppliers was not clearly defined.
CAPA verification/validation of corrective/preventive action
Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately and fully validated and approved according to established procedures.
Statistical methodology
Corrective and preventive action procedures addressing the use of appropriate statistical methodology to identify existing and potential causes of nonconforming product or other quality problems were not established, defined, documented, complete, and implemented.
Quality Audit/Reaudit - conducted
Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product were not established, defined, documented, complete, and implemented.
Evaluation of suppliers, contractors, etc., requirements
Adequate quality requirements that must be met by suppliers were not established, documented, and implemented.
Changes to correct/prevent quality problems
Changes in methods and procedures needed to correct and prevent identified quality problems are not implemented, recorded, and effective.
DMR device specifications
The device master record does not include or refer to the location of all and device specifications.
Documentation of disposition, justification, signature
There is incomplete documentation of the disposition of nonconforming product, justification for use of nonconforming product, and the signature of the individual authorizing the use of nonconforming product.
Analysis of data sources
Appropriate sources of quality data are not adequately analyzed to identify existing and potential causes of nonconforming product and other quality problems.
Analysis of data/reports from data sources
Not all data and reports from quality data sources are analyzed to identify existing and potential causes of nonconforming product and other quality problems.
Lack of or inadequate final acceptance procedures
Procedures for acceptance or rejection of finished device production runs, lots, or batches were not established, defined, documented, complete, and implemented.
General
Complaint handling procedures for receiving and evaluating complaints have not been defined and implemented.
Lack of or inadequate DHR procedures
Procedures have not been implemented to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
Documentation
Corrective and preventive action activities have not been documented, including analysis of sources of quality data, investigations of causes of nonconformities, and the actions needed to correct or prevent recurrence of nonconforming product and other quality problems.
Quality Audits - defined intervals
Quality audits did not verify that the quality system is effective in fulfilling your quality system objectives.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance or rejection of incoming product were not defined, documented, and complete.
Process control procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications were not established, defined, documented, and implemented.
Production monitoring and control
During production, component and device characteristics are not fully monitored and controlled.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.20Management responsibility2 of 2
- 21 CFR 820.40[Reserved]2 of 2
- 21 CFR 820.70Production and process controls2 of 2
- 21 CFR 820.75Process validation2 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 2
- 21 CFR 820.90Nonconforming product2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.181Device master record2 of 2
Show all 15
- 21 CFR 820.184Device history record2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 820.250Statistical techniques2 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.186Quality system record1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1527214 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1527214. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.