Torbot Group, Inc.
1367 Elmwood Ave, Cranston, Rhode Island, United States
FEI 1211665
FDA has 3 inspections on file for Torbot Group, Inc. in Cranston, Rhode Island (FEI 1211665), from 2009 to 2014. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 10 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Seven regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2009–2014
- 0 OAI · 3 VAI · 0 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Review and evaluation for investigation
Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Show 5 more citations
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Quality Audits - defined intervals
Quality audits were not performed to determine whether the quality system activities and results comply with quality system procedures.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Acceptance status
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria and maintained.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Documentation
Process validation activities and results have not been documented.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Equipment control activity documentation
Equipment calibrations, inspections, checks, and maintenance activities have not been documented.
Show 5 more citations
Lack of procedures, or not maintained
Document control procedures have not been established.
DHR - not or inadequately maintained
A device history record has not been maintained.
Event description
The events that led to reporting and/or conducting the correction or removal actions were not described.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Training records
Personnel training is not documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate DHR procedures
Procedures have not been established, documented, and implemented to ensure that device history records for each batch, lot, or unit are maintained to demonstrate that the device is manufactured in accordance with the device master record and the Quality Systems regulation.
General
Complaint handling procedures for receiving, reviewing, and evaluating complaints have not been established.
Training records
Employee training is not documented.
Procedures
The procedures addressing documentation of corrective and preventive action activities were not established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not established.
Show 1 more citation
Acceptance status
The acceptance status of product was not clearly identified.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.25Personnel3 of 3
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 820.184Device history record3 of 3
- 21 CFR 820.22Quality audit2 of 3
- 21 CFR 820.40[Reserved]2 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 3
- 21 CFR 820.86Acceptance status2 of 3
- 21 CFR 820.198Complaint files2 of 3
Show all 15
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 806.10Reports of corrections and removals1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.75Process validation1 of 3
- 21 CFR 820.90Nonconforming product1 of 3
- 21 CFR 820.181Device master record1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1211665 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1211665. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.