Health-Chem Acquisitions, LLC
3341 Sw 15th St, Pompano Beach, Florida, United States
FEI 1048532
FDA has 4 inspections on file for Health-Chem Acquisitions, LLC in Pompano Beach, Florida (FEI 1048532), from 2012 to 2019. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2012–2019
- 0 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design control - no procedures
Procedures for design control have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed in accordance with 21 CFR 820.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
DHR - not or inadequately maintained
A device history record has not been maintained.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of procedures, or not maintained
Document control procedures have not been established and maintained.
Show 6 more citations
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated and checked have not been adequately established.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of procedures, or not maintained
Document control procedures have not been established and maintained.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Show 7 more citations
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Distribution records
Distribution records were not maintained and do not include or refer to the location of required information.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate procedures for distribution
Procedures for control and distribution of finished devices have not been adequately established.
Production processes
Production processes were not developed, conducted, controlled, and monitored to ensure that a device conforms to its specifications.
Regulations FDA cited here
The 17 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 820.20Management responsibility2 of 3
- 21 CFR 820.25Personnel2 of 3
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.40[Reserved]2 of 3
- 21 CFR 820.50Purchasing controls2 of 3
- 21 CFR 820.75Process validation2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
Show all 17
- 21 CFR 820.22Quality audit1 of 3
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
- 21 CFR 820.90Nonconforming product1 of 3
- 21 CFR 820.160Distribution1 of 3
- 21 CFR 820.181Device master record1 of 3
- 21 CFR 820.184Device history record1 of 3
- 21 CFR 820.198Complaint files1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1048532 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1048532. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.