Sunnex, LLC

8001 Tower Point Dr, Charlotte, North Carolina, United States

FEI 1000583780

FDA has 6 inspections on file for Sunnex, LLC in Charlotte, North Carolina (FEI 1000583780), from 2010 to 2022. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 11 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

The most recent OAI inspection ended on Feb 18, 2015. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

14 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
6, 2010–2022
1 OAI · 4 VAI · 1 NAI
Inspections with 483 observations shown
4 of 6
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design verification - documentation

      The design verification results, including identification of the design, method(s), the date and the individual(s) performing the verification, were not documented in the design history file.

    2. Analyzing service report

      Service reports were not analyzed following appropriate statistical methods.

    3. Design validation - production units

      The design was not validated under defined operating conditions and using initial production units, lots or batches or their equivalents.

    4. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.

    5. Service reports

      Service reports do not include the required information.

    Show 6 more citations
    1. Documentation of validated process performance

      There is inadequate documentation of monitoring and control methods and data for a validated process.

    2. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

    3. Document review, approval by designated individual

      Documents were not reviewed and not approved by designated individual(s) prior to issuance .

    4. Design transfer - Lack of or inadequate procedures

      Procedures for design transfer have not been adequately established.

    5. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been established.

    6. Design plans - Lack of or inadequate

      Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design verification - documentation

      The design verification results, including, were not documented in the design history file.

    2. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    3. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been documented in the device history record.

    4. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been established.

    5. Service reports

      Service reports are not documented and do not include the required information.

    Show 7 more citations
    1. Document review, approval by designated individual

      Documents were not approved by designated individual(s) prior to issuance .

    2. Design transfer - Lack of or inadequate procedures

      Procedures for design transfer have not been established.

    3. Design validation - production units

      The design was not validated under defined operating conditions and using initial production units, lots or batches or their equivalents.

    4. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    5. Investigation of device failures

      Complaints involving the possible failure of a device, labeling and packaging to meet any of its specifications were not reviewed, evaluated and investigated where necessary.

    6. DHR documentation of label release {see also 820.184}

      The DHR does not include records of examination and release of device labeling, including date and signature of the examiner.

    7. Documentation

      Process validation activities and results have not been documented and approved.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Examination for accuracy

      Labeling was not sufficiently examined by a designated individual for accuracy including before release.

    2. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.

    3. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    4. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    5. Documentation of validated process performance

      There is no and inadequate documentation of monitoring and control methods and data, the date performed, the individual performing the process and the major equipment used for a validated process.

    Show 3 more citations
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    2. Document review, approval by designated individual

      Documents were not reviewed and not approved by designated individual(s) prior to issuance .

    3. Service reports

      Service reports do not include the required information.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices13 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    3. Training records

      Personnel training is not documented.

    4. DMR - not or inadequately maintained

      A device master record has not been maintained.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    Show 8 more citations
    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    2. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    3. Supplier notification of changes

      There is no agreement with suppliers to notify you of changes in the product or service.

    4. Lack of or inadequate DHR procedures

      Procedures for device history records have not been established.

    5. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers have not been established.

    6. Procedures for product review,disposition lack of/inadequate

      Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been established.

    7. Review and evaluation for investigation

      Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.

    8. Personnel

      Personnel do not have the necessary background, training, and experience to perform their jobs.

  6. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 6 inspections, with how many of them cited each.

Show all 14

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000583780 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000583780. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.