FDA has 7 inspections on file for Tenex Health, Inc. in Lake Forest, California (FEI 1000135560), from 2008 to 2021. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on Apr 11, 2013. FDA has issued 2 warning letters under this FEI, the latest on (CDRH). No court action or import refusal is filed under this FEI.
Eight regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 7, 2008–2021
- 2 OAI · 1 VAI · 4 NAI
- Inspections with 483 observations shown
- 4 of 7
- Form 483s posted by FDA
- None held for this FEI
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate and is incomplete.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Supplier oversight
The type and extent of control to be exercised over the product, services, and suppliers was not clearly defined.
Records of complaint investigation
Records of complaint investigations do not include required information.
Show 2 more citations
Design input - documentation
Design input requirements were not adequately documented.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Sampling plans
Sampling plans are not based on valid statistical rationale.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
Training records
Personnel training is not documented.
Show 8 more citations
Packaging
Device packaging and/or shipping containers are not designed and constructed to protect the device from alteration or damage during processing, storage, handling, and distribution.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
Documentation
Process validation activities have not been documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Software validation for automated processes
Software used as part of production and the quality system has not been validated for its intended use according to an established protocol.
Show 8 more citations
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.
Design validation - simulated testing
The design was not validated under actual or simulated use conditions.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Design input - documentation
Design input requirements were not adequately documented.
Design output - documentation
Design output was not adequately documented before release.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities were not implemented.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Nonconforming product, Lack of or inadequate procedures
Procedures were not followed for the control of products that do not conform to specifications.
Design changes - Lack of or Inadequate Procedures
Procedures were not followed for the validation or verification and review of design changes before their implementation.
General
Complaint handling procedures for receiving and evaluating complaints have not been implemented.
Show 1 more citation
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 7 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls4 of 4
- 21 CFR 820.198Complaint files4 of 4
- 21 CFR 820.75Process validation3 of 4
- 21 CFR 820.90Nonconforming product3 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 4
- 21 CFR 806.10Reports of corrections and removals2 of 4
- 21 CFR 820.40[Reserved]2 of 4
- 21 CFR 820.50Purchasing controls2 of 4
Show all 15
- 21 CFR 820.100Corrective and preventive action2 of 4
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 4
- 21 CFR 820.130Device packaging1 of 4
- 21 CFR 820.250Statistical techniques1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1000135560 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1000135560. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.