Helio USA Inc

816 Charcot Ave, San Jose, California, United States

FEI 1000135119

FDA has 3 inspections on file for Helio USA Inc in San Jose, California (FEI 1000135119), from 2011 to 2016. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 14 citations from this inspection: 13 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

No warning letter, court action or import refusal is filed under this FEI.

14 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2011–2016
0 OAI · 3 VAI · 0 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices14 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    2. Review and evaluation for investigation

      Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    4. Lack of or inadequate procedures for storage.

      Procedures for the control of storage areas and stock rooms have not been adequately established.

    5. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    Show 9 more citations
    1. QSR

      The quality system record has not been adequately maintained.

    2. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    3. Not approved or obsolete document retrieval

      Documents that were not approved and obsolete were observed at a location where they are being used.

    4. Info evaluated to determine if event was reportable

      The written MDR procedure does not include documentation and recordkeeping requirements for all information that was evaluated to determine if an event was reportable.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    6. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    7. Design transfer - Lack of or inadequate procedures

      Procedures for design transfer have not been established.

    8. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    9. Design control - no procedures

      Procedures for design control have not been established.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices22 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    2. Design validation - documentation

      The results of design validation, including identification of the design, method(s), the date, and the individual(s) performing validation, were not documented in the design history file.

    3. QSR

      The quality system record has not been adequately maintained

    4. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    5. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    Show 17 more citations
    1. Personnel

      Personnel do not have the necessary education, background, training, and experience to perform their jobs.

    2. Design control - no procedures

      Procedures for design control have not been established.

    3. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    4. Design verification - documentation

      The design verification results, including identification of the design, method(s), the date, and the individual(s) performing the verification, were not documented in the design history file.

    5. Design review - documentation

      The design review results, including identification of the design, the date, and the individual(s) performing the review, were not documented in the design history file.

    6. Lack of or inadequate procedures for storage.

      Procedures for the control of storage areas and stock rooms have not been adequately established.

    7. Documented evaluation

      The evaluation of potential suppliers, contractors, and consultants was not documented.

    8. Design transfer - Lack of or inadequate procedures

      Procedures for design transfer have not been established.

    9. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    10. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    11. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    12. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    13. Review and evaluation for investigation

      Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.

    14. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    15. Availability

      Required records are not maintained at a location that is reasonably accessible to responsible officials of the manufacturer and to employees of the FDA and were not made readily available for review and copying by the FDA.

    16. Lack of procedures, or not maintained

      Document control procedures have not been maintained.

    17. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Design control - no procedures

      Procedures for design control have not been established.

    4. Supplier notification of changes

      There is no agreement with suppliers and contractors to notify you of changes in the product or service.

    5. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    Show 4 more citations
    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    2. Training records

      Personnel training is not documented.

    3. Supplier oversight

      The type and extent of control to be exercised over the product, services, suppliers, and contractors was not clearly defined.

    4. Procedures not adequately established or maintained

      Document control procedures have not been adequately established and maintained.

Regulations FDA cited here

The 16 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Show all 16

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000135119 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000135119. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.