Skip to content

Parks Medical Electronics, Inc

19460 SW Shaw St, Aloha, Oregon, United States

FEI 3020115

FDA has 6 inspections on file for Parks Medical Electronics, Inc in Aloha, Oregon (FEI 3020115), from 2010 to 2023. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Seven regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
6, 2010–2023
0 OAI · 3 VAI · 3 NAI
Inspections with 483 observations shown
6 of 6
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Maintenance schedule, Lack of or inadequate schedule

      Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    3. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - documentation

      The results of design validation, including method(s), the date and the individual(s) performing validation , were not adequately documented in the design history file.

    2. Design verification - documentation

      The design verification results, including method(s) , were not adequately documented in the design history file.

    3. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated and maintained have not been adequately established.

    4. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    2. Software validation for automated processes

      Software used as part of production has not been validated for its intended use according to an established protocol.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Lack of procedures, or not maintained

      Document control procedures have not been established and maintained.

    4. Supplier oversight

      The type and extent of control to be exercised over services was not clearly defined.

    5. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.

    Show 5 more citations
    1. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with 21 CFR 820.

    2. Servicing - Lack of or inadequate procedures

      Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.

    3. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.

    4. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    5. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.

    2. Design validation - software validation documentation

      Results of the validation of the device software were not documented.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Complaints

      Complaint files are not adequately maintained.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    Show 3 more citations
    1. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    2. Design validation - Risk analysis

      Results of the design risk analysis were not adequately documented.

    3. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.

    2. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    3. Documentation

      Corrective and preventive action activities and/or results have not been documented.

Regulations FDA cited here

The 13 sections of 21 CFR in FDA’s published citations for this facility’s 6 inspections, with how many of them cited each.

Show all 13

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3020115 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3020115. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.