OSKO, Inc.
8085 Nw 90th St, Medley, Florida, United States
FEI 3010666400
FDA has 2 inspections on file for OSKO, Inc. in Medley, Florida (FEI 3010666400), from 2017 to 2019. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 19 citations from this inspection. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 2, 2017–2019
- 0 OAI · 1 VAI · 1 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices19 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Failure to submit, distinct marking
You did not submit a required product report prior to the introduction of the product into commerce.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been established.
Production processes
Production processes were not developed, conducted, controlled and monitored to ensure that a device conforms to its specifications.
Design review - Lack of or inadequate procedures
Procedures for design review have not been established.
Show 14 more citations
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Compliance measurement
Compliance with the aluminum equivalence limits was not determined by x-ray measurements made at a potential of 100 kilovolts peak and with an x-ray beam that has a half-value layer of 2.7 millimeters of aluminum.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been established.
Software validation for automated processes
Software used as part of production and the quality system has not been validated for its intended use according to an established protocol.
Compliance measurements--diagnostic source assembly
Compliance for radiation leakage from the diagnostic source assembly was not determined by measurements averaged over an area of 100 square centimeters with no linear dimension greater than 20 centimeters.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Methods, devices, procedures for testing
You have not records of the used in tests for electronic product radiation safety.
Lack of or inadequate instructions
Installation instructions have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
FDA not notified of defect or noncompliance
You did not notify the Secretary, in accordance with 21 CFR 1003.20, of an electronic product which failed to comply with an applicable Federal standard which you produced, assembled and distributed.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Design input - documentation
Design input requirements were not documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Quality audits - auditor independence
Individuals who conduct quality audits have direct responsibility for the matters being audited.
Independent mA selection
For equipment having independent selection of x-ray tube current, the average ratios of exposure to the indicated milliampere-seconds product obtained at any two consecutive tube current settings differed by more than 0.10 times their sum.
Software validation for automated processes
Software used as part of production and the quality system has not been validated for its intended use according to an established protocol.
Methods, devices, procedures for testing
You have not established and maintained records of the methods, devices and procedures used in tests for electronic product radiation safety.
Show 4 more citations
Procedures not adequately established or maintained
Document control procedures have not been adequately maintained.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Half-value layer--amount of filtration
With reference to Table 1 appearing in Title 21 CFR 1020.30(m), the half-value layer (HVL) of the useful beam for a given x-ray tube potential was less than the appropriate value shown in table 1.
Regulations FDA cited here
The 17 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.70Production and process controls2 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 1002.30Records to be maintained by manufacturers2 of 2
- 21 CFR 1020.30Diagnostic x-ray systems and their major components2 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.25Personnel1 of 2
- 21 CFR 820.30Design controls1 of 2
Show all 17
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.170Installation1 of 2
- 21 CFR 1002.10Product reports1 of 2
- 21 CFR 1003.10Discovery of defect or failure of compliance by manufacturer; notice requirements1 of 2
- 21 CFR 1020.31Radiographic equipment1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3010666400 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3010666400. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.