Innovative Health Technologies, LLC
2355 S 1070 W Ste D, Salt Lake City, Utah, United States
FEI 3010078139
FDA has 3 inspections on file for Innovative Health Technologies, LLC in Salt Lake City, Utah (FEI 3010078139), from 2011 to 2018. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.30(g) and 21 CFR 820.100(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2011–2018
- 0 OAI · 3 VAI · 0 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - software validation documentation
Results of the validation of the device software were not adequately documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation of software validation
Software validation activities and results for computers or automated data processing systems used as part of production have not been adequately documented.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.
Acceptable supplier records, inadequate records
Records of acceptable suppliers have not been adequately established.
Design validation - software validation documentation
Results of the validation of the device software were not adequately documented.
Equipment suitability & capability
Certain inspection, measuring, and test equipment is not suitable for its intended purposes.
Show 1 more citation
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality System Procedures
Quality system procedures and instructions have not been established.
DMR - not or inadequately maintained
A device master record has not been maintained.
Product rework procedures, Lack of or inadequate procedures
Procedures for rework of nonconforming product have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Supplier notification of changes
There is no agreement with suppliers and contractors to notify you of changes in the product or service.
Show 11 more citations
Training records
Personnel training is not documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities have not been established.
Design validation - software validation not performed
Validation of device software was not performed.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential suppliers and contractors were not evaluated and selected based on their ability to meet specified requirements.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls3 of 3
- 21 CFR 820.50Purchasing controls2 of 3
- 21 CFR 820.100Corrective and preventive action2 of 3
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.40[Reserved]1 of 3
Show all 15
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
- 21 CFR 820.90Nonconforming product1 of 3
- 21 CFR 820.181Device master record1 of 3
- 21 CFR 820.184Device history record1 of 3
- 21 CFR 820.200Servicing1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3010078139 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3010078139. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.