W & R Investments LLC d.b.a. Laser Engineering
113 Cedar St Ste S5, Milford, Massachusetts, United States
FEI 3009593803
FDA has 4 inspections on file for W & R Investments LLC d.b.a. Laser Engineering in Milford, Massachusetts (FEI 3009593803), from 2013 to 2022. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published one citation from this inspection, under 21 CFR Part 820, the device quality system rules. It is listed below.
The most recent OAI inspection ended on Apr 13, 2016. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.30(g), 21 CFR 820.22 and 21 CFR 820.90(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2013–2022
- 1 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety Act2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - production units
The design was not validated using initial production units, lots or batches or their equivalents.
Design validation acceptance criteria
Acceptance criteria were not established prior to the performance of validation activities.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a specification , method , process and procedure have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been adequately established.
Show 2 more citations
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Records of complaint investigation
Records of complaint investigations do not include required information.
Remedial action - documentation
Evaluations of out-of-calibration equipment and remedial actions taken were not documented.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Procedures not adequately established or maintained
Document control procedures have not been adequately established and maintained.
Show 4 more citations
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Lack of or inadequate In-process acceptance procedures
Procedures for the control of in-process product have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.22Quality audit2 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.70Production and process controls2 of 4
- 21 CFR 820.90Nonconforming product2 of 4
- 21 CFR 820.198Complaint files2 of 4
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
Show all 13
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 4
- 21 CFR 820.100Corrective and preventive action1 of 4
- 21 CFR 820.181Device master record1 of 4
- 21 CFR 820.184Device history record1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3009593803 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3009593803. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.