FDA has 5 inspections on file for BURTON MEDICAL in Addison, Illinois (FEI 3009542956), from 2013 to 2026. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA posted the Form 483 from that inspection. FDA published one citation from this inspection, under 21 CFR Part 830, the unique device identification rules. It is listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

One regulation was cited at more than one inspection: 21 CFR 820.100(a).

Sources:FDA firm profile ↗Form 483 PDF ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
5, 2013–2026
0 OAI · 3 VAI · 2 NAI
Inspections with 483 observations shown
4 of 5
Form 483s posted by FDA
↗
Requested under FOIA
2 requests, latest May 19, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesPosted 483 · 1 citation

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Devices subject to device identification GUDID data submission requirements.

      The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    2. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    3. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Submission of Supplement to Individual Report Within One Month

      A supplemental report was not submitted to FDA within one month following receipt of information that was not provided when the initial report was submitted.

    2. Records of MDR Investigation

      Investigation records of MDR reportable complaints do not include required information.

    3. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for the timely and effective communication of events that may be subject to medical device reporting requirements.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

    2. Personnel

      Personnel do not have the necessary education, background, training, and experience to perform their jobs.

    3. Quality audit - audit report review

      Reports of results of quality audits are not reviewed by management having responsibility for the matters audited.

    4. Product rework procedures, Lack of or inadequate procedures

      Procedures for rework of nonconforming product have not been established.

    5. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been established.

    Show 6 more citations
    1. Quality plan

      A quality plan has not been adequately established.

    2. Analyzing service report

      Service reports were not analyzed following appropriate statistical methods.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Document review, approval by designated individual

      Documents were not reviewed and not approved by designated individual(s) prior to issuance .

    5. Design history file

      The design history file was not established, does not demonstrate that the design was developed following the approved design plan, and does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    6. Product rework adverse effects {see also 820.184}

      Documentation of rework and reevaluation activities does not include a determination of whether there has been any adverse effect from rework upon the product.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on May 19, 2026. Each one asked for a Form 483.

  • Inspection ending Feb 20, 2026NAI

    2 requests. In the inspection history above. FDA has posted its Form 483.

    1. May 19, 2026FDA-FOIA-2026-5148Form 483· 1 of 4 sitesRecords released
    2. Apr 17, 2026FDA-FOIA-2026-3890Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.

  • 21 CFR 820.198Complaint files3 of 4
  • 21 CFR 820.30Design controls2 of 4
  • 21 CFR 820.100Corrective and preventive action2 of 4
  • 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 4
  • 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 4
  • 21 CFR 803.56If I am a manufacturer, in what circumstances must I submit a supplemental or followup report and what are the requirements for such reports?1 of 4
  • 21 CFR 820.20Management responsibility1 of 4
  • 21 CFR 820.22Quality audit1 of 4
Show all 14

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3009542956 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3009542956. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.