Reflexonic, LLC
8180 Greensboro Dr Ste 300, Mc Lean, Virginia, United States
FEI 3009129589
FDA has 2 inspections on file for Reflexonic, LLC in Mc Lean, Virginia (FEI 3009129589), from 2016 to 2018. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 14 citations from this inspection: 13 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below. One FOIA request asked FDA for the 483 from a later inspection, ending Sep 24, 2025, that FDA's Data Dashboard doesn't list yet.
No warning letter, court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2016–2018
- 0 OAI · 2 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Nov 13, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Software validation for automated processes
Software used as part of production has not been adequately validated for its intended use according to an established protocol.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 9 more citations
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated and investigated where necessary.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Training records
Personnel training is not documented.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design input - Lack of or inadequate procedures
Procedures for design input have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Show 4 more citations
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Distribution records
Distribution records were not maintained.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Nov 13, 2025. It asked for a Form 483.
Inspection ending Sep 24, 2025
1 request. Not listed in FDA’s Data Dashboard yet. FDA released records in answer to a request for its Form 483 on Jul 20, 2026. FDAgent holds no copy that FDA posted.
- Nov 13, 2025FDA-FOIA-2025-12905Form 483· 1 of 5 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.25Personnel2 of 2
- 21 CFR 820.30Design controls2 of 2
- 21 CFR 820.90Nonconforming product2 of 2
- 21 CFR 820.100Corrective and preventive action2 of 2
- 21 CFR 820.198Complaint files2 of 2
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 2
- 21 CFR 820.40[Reserved]1 of 2
- 21 CFR 820.50Purchasing controls1 of 2
Show all 13
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
- 21 CFR 820.160Distribution1 of 2
- 21 CFR 820.181Device master record1 of 2
- 21 CFR 820.200Servicing1 of 2
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3009129589 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3009129589. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.