Anatomage, Inc.
3350 Scott Blvd Bldg 29, Santa Clara, California, United States
FEI 3008272529
FDA has one inspection on file for Anatomage, Inc. in Santa Clara, California (FEI 3008272529). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 14 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2014
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices14 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Supplier notification of changes
There is no agreement with suppliers, contractors, and consultants to notify you of changes in the product or service.
Identification procedures, Lack of or inadequate procedures
Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.
Equipment control activity documentation
Equipment calibrations have not been documented.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Equipment Installation, Placement, Specified Requirements
The appropriate design, construction, placement, and installation of manufacturing equipment have not been ensured.
Show 9 more citations
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they are being used.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Documentation
Process validation activities have not been adequately documented.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
Review and evaluation for investigation
Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.40[Reserved]1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.60Identification1 of 1
- 21 CFR 820.70Production and process controls1 of 1
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 1
- 21 CFR 820.75Process validation1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
Show all 12
- 21 CFR 820.90Nonconforming product1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.184Device history record1 of 1
- 21 CFR 820.198Complaint files1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008272529 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008272529. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.