PACKAGE SOLUTIONS LLC
2131 Blount Rd, Pompano Beach, Florida, United States
FEI 3006559549
FDA has 2 inspections on file for PACKAGE SOLUTIONS LLC in Pompano Beach, Florida (FEI 3006559549), from 2013 to 2024. The latest ended on and was classified VAI, voluntary action indicated. FDA has also posted a Form 483 dated Feb 10, 2026 that FDAgent has not matched to an inspection on the Data Dashboard.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 10 citations from this inspection: 8 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 801, the device labeling rules, and 1 under 21 CFR Part 830, the unique device identification rules. They are listed below. One FOIA request has asked FDA for this inspection's 483, and two more for the 483 from a later inspection, ending Feb 10, 2026, that FDA's Data Dashboard doesn't list yet.
The most recent OAI inspection ended on Oct 3, 2013. 1 import refusal is filed under this FEI, on . FDA’s Import Alert 66-41 Red List names this firm at this address. FDA printed no FEI on the entry. No warning letter or court action is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 2, 2013–2024
- 1 OAI · 1 VAI · 0 NAI
- Inspections with 483 observations shown
- 2 of 2
- 1 posted 483 not matched to an inspection
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 3 requests, latest May 19, 2026
- Warning letters
- None on file
- Import refusals
- 1 refusal
- Import alerts
- 66-41Red List
- Matched on FDA’s firm name and street address
Inspection history
2 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC
- Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers, contractors and consultants have not been adequately established.
Quality Audits - defined intervals
Quality audits were not performed at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.
Complaints
Complaint files are not adequately maintained.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Show 5 more citations
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality System Procedures
Quality system procedures and instructions have not been established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been established.
Written warehousing procedures established/followed
Procedures describing the warehousing of drug products are not established.
Complaint Handling Procedure
Procedures describing the handling of written and oral complaints related to drug products are not written or followed.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Show 2 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on May 19, 2026. Each one asked for a Form 483.
Inspection ending Feb 10, 2026
2 requests. Not listed in FDA’s Data Dashboard yet. FDA has posted its Form 483.
- May 19, 2026FDA-FOIA-2026-5137Form 483· 1 of 4 sitesRecords released
- Apr 13, 2026FDA-FOIA-2026-3674Form 483· 1 of 5 sitesRecords released
Inspection ending Jun 18, 2024VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 21, 2024FDA-FOIA-2024-9301Form 483· 1 of 2 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for this facility’s 2 inspections, with how many of them cited each.
- 21 CFR 820.20Management responsibility2 of 2
- 21 CFR 820.50Purchasing controls2 of 2
- 21 CFR 211.142Warehousing procedures1 of 2
- 21 CFR 211.198Complaint files1 of 2
- 21 CFR 801.20Label to bear a unique device identifier1 of 2
- 21 CFR 820.22Quality audit1 of 2
- 21 CFR 820.30Design controls1 of 2
- 21 CFR 820.40[Reserved]1 of 2
Show all 15
- 21 CFR 820.70Production and process controls1 of 2
- 21 CFR 820.75Process validation1 of 2
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 2
- 21 CFR 820.90Nonconforming product1 of 2
- 21 CFR 820.100Corrective and preventive action1 of 2
- 21 CFR 820.198Complaint files1 of 2
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 2
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3006559549, one row per refused product line.
1 refusal on 2018-01-23 · charged NOT LISTED on 1, FRNMFGREG on 1 · no FDA sample analysis
FDA firm profile- 2018-01-23
REGULATOR N.E.C.65PAJ99
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
UNAPPROVED 505(a), 801(a)(3); UNAPPROVED NEW DRUG
Division of Southeast ImportsUPS-6919509-8/21/1
Import alerts
FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.
- 66-41Red ListDetention Without Physical Examination of Unapproved New Drugs Promoted In The U.S.Alert on fda.gov
Date published 2018-04-26
- FDA’s entry
- Package Solutions LLC · 2131 Blount Rd, Pompano Beach, FL 33069-5112 UNITED STATES
- FDA’s inspection record for FEI 3006559549
- PACKAGE SOLUTIONS LLC · 2131 Blount Rd, Pompano Beach, United States
Matched on exact firm name, street address and country. FDA printed no FEI on this entry.
4 products
- 54 A - - 99Vitamin, N.E.C.
- 54 Y - - 99Vitamin, Mineral, Proteins and Unconventional Dietary Specialities For Humans and Animals, N.E.C.
- 65 P - - 99Regulator N.E.C.
Show 1 more products
- 66 V - - 99Miscellaneous Patent Medicines, Etc.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006559549 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006559549. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.