Arnold Tuber Industries, LLC
10 Centre Dr Ste 1, Orchard Park, New York, United States
FEI 3003988575
FDA has 8 inspections on file for Arnold Tuber Industries, LLC in Orchard Park, New York (FEI 3003988575), from 2009 to 2025. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
11 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 8, 2009–2025
- 0 OAI · 6 VAI · 2 NAI
- Inspections with 483 observations shown
- 8 of 8
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Mar 21, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration procedures - content
Calibration procedures do not include provisions for remedial action.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Software validation for automated processes
Software used as part of production and the quality system has not been validated for its intended use according to an established protocol.
Lack of or inadequate In-process acceptance procedures
Procedures for the acceptance of in-process product have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Records of complaint investigation
Records of complaint investigations do not include required information.
Lack of or inadequate In-process acceptance procedures
Procedures for the acceptance and control of in-process product have not been established.
Software validation for automated processes
Software used as part of production has not been validated for its intended use according to an established protocol.
Show 2 more citations
Sampling plans
Sampling plans are not based on valid statistical rationale.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Lack of procedures, or not maintained
Document control procedures have not been maintained.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Complaints
Complaint files are not adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Lack of Written MDR Procedures
Written MDR procedures have not been developed, maintained, and implemented.
Design control - no procedures
Procedures for design control have not been established.
Documentation of software validation
Software validation activities and results for computers or automated data processing systems used as part of production and the quality system have not been documented.
Show 2 more citations
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Production processes
Production processes were not conducted, controlled, and monitored to ensure that a device conforms to its specifications.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices23 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design control - no procedures
Procedures to control the design process of the device were not defined, documented, and implemented.
Management review - defined interval, sufficient frequency
Management reviews were not conducted.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Documentation - specific items
Documentation of process validation activities and results does not include the date and signature of the individuals approving the validation and the major equipment validated.
Documentation
Corrective and preventive action activities have not been documented, including analysis of sources of quality data, the verification or validation of corrective actions, dissemination of information about quality problems or nonconforming product to responsible parties, and submission of information on quality problems and corrective and preventive actions for management review.
Show 18 more citations
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures were not defined and documented for monitoring and control of process parameters for validated processes.
Quality System Procedures
Quality system procedures were not complete and implemented.
Resources
Adequate resources have not been provided for performing management activities, work, assessment activities, and audits.
Quality audits - Lack of or inadequate procedures
Procedures for conducting quality audits were not established.
Management responsibility
Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.
Quality policy and objectives
The quality policy and objectives were not defined, documented, and implemented.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
DMR device specifications
The device master record does not include or refer to the location of all specifications.
Lack of or inadequate final acceptance procedures
Procedures for acceptance or rejection of finished device production runs, lots, or batches were not defined and documented.
Design history file
The design history file was not established or maintained.
Management representative
No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
Procedures
The procedures addressing documentation of corrective and preventive action activities were not complete and implemented.
General
Complaint handling procedures for complaints have not been completed and implemented.
Documentation of validated process performance
There is no documentation of monitoring and control methods and data, the date performed, and the individual performing the process for a validated process.
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
Training
Personnel are not adequately trained to perform management activities, work, assessment activities, and audits.
Activity documentation
Maintenance activities, including the date and individuals performing those activities, have not been documented.
Quality Audit/Reaudit - conducted
Quality audits were not conducted to verify that the quality system is effective in fulfilling your quality system objectives.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Mar 21, 2025. It asked for a Form 483.
Inspection ending Feb 20, 2025VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Mar 21, 2025FDA-FOIA-2025-2632Form 483· 1 of 5 sitesClosed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls4 of 8
- 21 CFR 820.70Production and process controls4 of 8
- 21 CFR 820.100Corrective and preventive action4 of 8
- 21 CFR 820.181Device master record4 of 8
- 21 CFR 820.198Complaint files4 of 8
- 21 CFR 820.75Process validation3 of 8
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 of 8
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 8
Show all 15
- 21 CFR 820.20Management responsibility2 of 8
- 21 CFR 820.22Quality audit2 of 8
- 21 CFR 820.184Device history record2 of 8
- 21 CFR 820.40[Reserved]1 of 8
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 8
- 21 CFR 820.90Nonconforming product1 of 8
- 21 CFR 820.250Statistical techniques1 of 8
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003988575 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003988575. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.