FDA has 3 inspections on file for Galloway Technologies, LLC in Saratoga Springs, Utah (FEI 3003843036), from 2011 to 2013. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 11 citations from this inspection: 10 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.

The most recent OAI inspection ended on Mar 2, 2011. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

11 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2011–2013
1 OAI · 2 VAI · 0 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Service reports

      Service reports do not include the required information.

    2. DMR device specifications

      The device master record does not include or refer to the location of device software specifications.

    3. Training records

      Personnel training is not documented.

    4. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    Show 6 more citations
    1. Review and evaluation for investigation

      Not all complaints have been adequately reviewed and evaluated to determine whether an investigation is necessary.

    2. Approval, inadequate purchasing data

      Purchasing data that clearly describe or reference specified requirements for purchased or otherwise received product and services have not been adequately established.

    3. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    4. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    5. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    6. Design validation - software validation not performed

      Validation of device software was not performed.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Service reports

      Service reports are not documented.

    2. Review and evaluation for investigation

      Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.

    3. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    4. Training records

      Personnel training is not documented.

    5. DHR - not or inadequately maintained

      A device history record has not been maintained.

    Show 7 more citations
    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    3. DMR - not or inadequately maintained

      A device master record has not been maintained.

    4. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    5. Approval, inadequate purchasing data

      Purchasing data that clearly describe or reference specified requirements for purchased or otherwise received product and services have not been adequately established.

    6. Design validation - software validation not performed

      Validation of device software was not performed.

    7. Supplier notification of changes

      There is no agreement with suppliers, contractors, and consultants to notify you of changes in the product or service.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices20 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality System Procedures

      Quality system procedures and instructions have not been established.

    2. Design control - no procedures

      Procedures for design control have not been established.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    4. DMR - not or inadequately maintained

      A device master record has not been maintained.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    Show 15 more citations
    1. DHR - not or inadequately maintained

      A device history record has not been maintained.

    2. Design validation - software validation not performed

      Validation of device software was not performed.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    4. Supplier notification of changes

      There is no agreement with suppliers and contractors to notify you of changes in the product or service.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    6. Service reports

      Service reports are not documented.

    7. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    8. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential suppliers and contractors were not evaluated and selected based on their ability to meet specified requirements.

    9. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    10. Design Validation - Risk analysis not performed/inadequate

      Risk analysis was not performed.

    11. Incoming acceptance records, documentation

      Acceptance or rejection of incoming product was not documented.

    12. Lack of procedures, or not maintained

      Document control procedures have not been established.

    13. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.

    14. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    15. Training records

      Personnel training is not documented.

Regulations FDA cited here

The 15 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Show all 15

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003843036 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003843036. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.