Gulf Fiberoptics, Inc.
448 Commerce Blvd, Oldsmar, Florida, United States
FEI 3003768921
FDA has 8 inspections on file for Gulf Fiberoptics, Inc. in Oldsmar, Florida (FEI 3003768921), from 2010 to 2023. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued.
The most recent OAI inspection ended on May 3, 2011. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
11 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 8, 2010–2023
- 1 OAI · 6 VAI · 1 NAI
- Inspections with 483 observations shown
- 7 of 8
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
8 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Documentation
In-process inspections, tests, or other verification activities and approvals were not documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Documentation
Acceptance activities were not adequately documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Documentation
Acceptance activities were not adequately documented.
Design input - documentation
Design input requirements were not documented.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers, contractors, and consultants have not been adequately established.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a specification, method, process, and procedure have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of procedures, or not maintained
Document control procedures have not been established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Documentation
Acceptance activities were not documented.
Personnel
Personnel do not have the necessary background, training, and experience to perform their jobs.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 5 more citations
Documentation
Process validation activities have not been adequately documented.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of System for Event Evaluations
The written MDR Procedure does not include an internal system which provides for the timely and effective identification, communication, and evaluation of events that may be subject to medical device reporting requirements.
Lack of System for Determining MDR Events
The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.
Design plans - Lack of or inadequate
Design plans that describe or reference the design and development activities and define responsibility for implementation have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Procedures not adequately established or maintained
Document control procedures have not been adequately established and maintained.
Show 6 more citations
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Documentation
Acceptance activities were not documented and adequately documented.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Records of complaint investigation
Records of complaint investigations do not include required information.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audits - defined intervals
Quality audits did not verify that the quality system is effective in fulfilling your quality system objectives.
Incoming inspection, testing, verification
Incoming product was not adequately inspected or tested to verify conformance to specifications.
Design plans - Lack of or inadequate
The design plan does not describe the design and development activities and define responsibility for implementation of design and development activities.
Lack of or inadequate final acceptance procedures
Procedures for acceptance or rejection of finished device production runs, lots, or batches were not established.
Management representative
No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
Show 6 more citations
Acceptance records
The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.
Nonconforming product, Lack of or inadequate procedures
Procedures were not followed for the control of products that do not conform to specifications.
Documentation of validated process performance
There is no documentation of monitoring and control methods and data, the date performed, the individual performing the process, and the major equipment used for a validated process.
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
Quality audits - Lack of or inadequate procedures
Procedures for conducting quality audits were not defined and complete.
Review and evaluation for investigation
Not all complaints have been reviewed and evaluated to determine whether an investigation is necessary.
Regulations FDA cited here
The 15 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 8 inspections, with how many of them cited each.
- 21 CFR 820.80Receiving, in-process, and finished device acceptance6 of 7
- 21 CFR 820.30Design controls4 of 7
- 21 CFR 820.75Process validation4 of 7
- 21 CFR 820.90Nonconforming product4 of 7
- 21 CFR 820.100Corrective and preventive action4 of 7
- 21 CFR 820.70Production and process controls3 of 7
- 21 CFR 820.20Management responsibility2 of 7
- 21 CFR 820.22Quality audit2 of 7
Show all 15
- 21 CFR 820.40[Reserved]2 of 7
- 21 CFR 820.184Device history record2 of 7
- 21 CFR 820.198Complaint files2 of 7
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 7
- 21 CFR 820.25Personnel1 of 7
- 21 CFR 820.50Purchasing controls1 of 7
- 21 CFR 820.181Device master record1 of 7
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003768921 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003768921. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.