North American Medical Corp. (NAM)

1649 Sands Pl Se Ste A, Marietta, Georgia, United States

FEI 3003610317

FDA has 4 inspections on file for North American Medical Corp. (NAM) in Marietta, Georgia (FEI 3003610317), from 2009 to 2022. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

No warning letter, court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.25(a), 21 CFR 820.25(b) and 21 CFR 820.90(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
4, 2009–2022
0 OAI · 3 VAI · 1 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices13 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Personnel

      Personnel do not have the necessary training to perform their jobs.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    5. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    Show 8 more citations
    1. Acceptable supplier records, inadequate records

      Records of acceptable suppliers have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Complaints

      Complaint files are not adequately maintained. Electronic records are used, but they do not meet audit trail requirements to ensure that they are trustworthy, reliable and generally equivalent to paper records.

    4. Lack of or inadequate procedures for labeling

      Procedures to control labeling activities have not been adequately established.

    5. Documentation

      In-process inspections, tests, or other verification activities and approvals were not documented.

    6. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    7. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    8. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    2. Personnel

      Personnel do not have the necessary background, training, and experience to perform their jobs.

    3. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    4. Maintained

      Records of complaint investigations are not maintained by the formally designated unit.

    5. Production processes

      Production processes were not controlled and monitored to ensure that a device conforms to its specifications.

    Show 6 more citations
    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not evaluated and investigated where necessary.

    2. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    3. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    4. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    5. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been fully documented in the device history record.

    6. Documentation

      Acceptance activities were not documented and maintained as part of the device history record.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate final acceptance procedures

      Procedures for acceptance or rejection of finished device production runs, lots, or batches were not established and defined.

    2. CAPA verification/validation of corrective/preventive action

      Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.

    3. Incoming inspection, testing, verification

      Incoming product was not adequately inspected or tested to verify conformance to specifications.

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

Show all 14

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003610317 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003610317. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.