Veolia Water Treatment & Solutions
1217 W Corporate Dr, Arlington, Texas, United States
FEI 3003550196
FDA has 4 inspections on file for Veolia Water Treatment & Solutions in Arlington, Texas (FEI 3003550196), from 2009 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 3 citations from this inspection: 1 under 21 CFR Part 801, the device labeling rules, 1 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 830, the unique device identification rules. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Feb 26, 2009. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
One regulation was cited at more than one inspection: 21 CFR 820.90(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2009–2025
- 1 OAI · 1 VAI · 2 NAI
- Inspections with 483 observations shown
- 3 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest FY2023
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Devices subject to device identification GUDID data submission requirements.
The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.
Devices that must be directly marked with a unique device identifier
The medical device(s) that is intended to be used more than once and intended to be reprocessed before each use was not directly marked with a unique device identifier (UDI).
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices17 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation- Lack of or inadequate procedures
Procedures for validating the device design were not established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Design verification - Lack of or inadequate procedures
Design verification did not confirm that the design output meets the design input requirements.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance or rejection of incoming product were not established, defined, documented, and implemented.
Design output - documentation
Design output was not documented before release.
Show 12 more citations
Design verification procedures
Procedures for verifying that design output meets design input were not established.
Management representative
No management representative had been appointed to ensure that quality system requirements are met, and to report to management on the performance of the quality system.
Lack of procedures, or not maintained
Document control procedures were not complete and implemented.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not complete and implemented.
Quality audits - Lack of or inadequate procedures
Procedures for conducting quality audits were not established and complete.
Nonconforming product, Lack of or inadequate procedures
Procedures were not complete for the control of products that do not conform to specifications.
Design history file
The design history file was not established or maintained.
Design output procedures
Procedures for defining and documenting design output in terms that allow an adequate evaluation of conformance to design input requirements were not established.
Design control - no procedures
Procedures to control the design process of the device were not established.
Design input procedures - appropriateness
Procedures to ensure that a device's design input requirements are appropriate and address its intended use, including user/patient needs, were not established, defined, and documented.
Installer's documentation of results
The person installing the device did not document complete inspection and test results to demonstrate proper installation.
Design review procedures - personnel
Procedures were not established, defined, and documented for ensuring that participants at each design review include an individual who does not have direct responsibility for the design stage being reviewed.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on FY2023. Each one asked for a Form 483.
Inspection ending Oct 17, 2025NAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Mar 17, 2026. FDAgent holds no copy that FDA posted.
- Dec 18, 2025FDA-FOIA-2025-14176Form 483· 1 of 5 sitesRecords released
Inspection ending Sep 16, 2022VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- FY2023FDA-FOIA-2023-84Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.80Receiving, in-process, and finished device acceptance2 of 3
- 21 CFR 820.90Nonconforming product2 of 3
- 21 CFR 820.198Complaint files2 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.22Quality audit1 of 3
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.40[Reserved]1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
Show all 13
- 21 CFR 820.70Production and process controls1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
- 21 CFR 820.170Installation1 of 3
- 21 CFR 830.300Devices subject to device identification data submission requirements1 of 3
- 21 CFR 801.45Devices that must be directly marked with a unique device identifier1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003550196 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003550196. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.