21 CFR 801.45: Devices that must be directly marked with a unique device identifier
FDA's inspection data lists 3 Form 483 citations of 21 CFR 801.45 at 3 facilities, from November 2023 to October 22, 2025.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 3
- At 3 inspections
- Facilities cited
- 3
- Distinct FEIs
- Most recent citation
- Oct 22, 2025
- Sodium Systems LLC
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 801.45 from the Electronic Code of Federal Regulations.
§ 801.45 Devices that must be directly marked with a unique device identifier.
(a) In general. A device that must bear a unique device identifier (UDI) on its label must also bear a permanent marking providing the UDI on the device itself if the device is intended to be used more than once and intended to be reprocessed before each use.
(b) UDI for direct marking. The UDI provided through a direct marking on a device may be:
(1) Identical to the UDI that appears on the label of the device, or
(2) A different UDI used to distinguish the unpackaged device from any device package containing the device.
(c) Form of a UDI when provided as a direct marking. When a device must bear a UDI as a direct marking, the UDI may be provided through either or both of the following:
(1) Easily readable plain-text;
(2) Automatic identification and data capture (AIDC) technology, or any alternative technology, that will provide the UDI of the device on demand.
(d) Exceptions. The requirement of paragraph (a) of this section shall not apply to any device that meets any of the following criteria:
(1) Any type of direct marking would interfere with the safety or effectiveness of the device;
(2) The device cannot be directly marked because it is not technologically feasible;
(3) The device is a single-use device and is subjected to additional processing and manufacturing for the purpose of an additional single use.
(4) The device has been previously marked under paragraph (a) of this section.
(e) Exception to be noted in design and development files. A labeler that decides to make use of an exception under paragraph (d) of this section must document the basis of that decision in the design and development files required by § 820.10(c) of this chapter.
21 CFR Part 801: Labeling, Subpart BRead on eCFR
Citations by year
Form 483 citations of 21 CFR 801.45 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 2 in 2025
| Year | Form 483 citations |
|---|---|
| 2023 | 1 |
| 2025 | 2 |
How FDA words this citation
FDA's short descriptions and citation text on these 3 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Devices that must be directly marked with a unique device identifier3 citations
Citation text
- “The medical device(s) that is intended to be used more than once and intended to be reprocessed before each use was not directly marked with a unique device identifier (UDI).”3 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 801.45, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Sodium Systems LLC2025-10-22 · FEI 3014529181Devices that must be directly marked with a unique device identifier
- Veolia Water Treatment & Solutions2025-10-17 · FEI 3003550196Devices that must be directly marked with a unique device identifier
- Princeton Medical Group Inc2023-11-21 · FEI 1000111204Devices that must be directly marked with a unique device identifier
Often cited at the same inspection
Of the 3 inspections where FDA cited 21 CFR 801.45, the share where it also cited each of these sections.
- 21 CFR 820.198Complaint files2 (67%)
- 21 CFR 820.100Corrective and preventive action1 (33%)
- 21 CFR 820.184Device history record1 (33%)
- 21 CFR 820.30Design controls1 (33%)
- 21 CFR 820.50Purchasing controls1 (33%)
- 21 CFR 830.300Devices subject to device identification data submission requirements1 (33%)
Ask FDAgent about 21 CFR 801.45
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 801.45; FDAgent quotes that sentence exactly.