Sodium Systems LLC
540 Avis Dr Ste H, Ann Arbor, Michigan, United States
FEI 3014529181
FDA has one inspection on file for Sodium Systems LLC in Ann Arbor, Michigan (FEI 3014529181). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 5 citations from this inspection: 4 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 801, the device labeling rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2025
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 2 requests, latest Jan 21, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Records of complaint investigation - UDI
Records of complaint investigations do not include required information, including any unique device identifier (UDI) or universal product code (UPC).
Devices that must be directly marked with a unique device identifier
The medical device(s) that is intended to be used more than once and intended to be reprocessed before each use was not directly marked with a unique device identifier (UDI).
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Two requests to FDA under the Freedom of Information Act named this site, most recently on Jan 21, 2026. Each one asked for a Form 483.
Inspection ending Oct 22, 2025VAI
2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Mar 3, 2026. FDAgent holds no copy that FDA posted.
- Jan 21, 2026FDA-FOIA-2026-671Form 483· 1 of 5 sitesOpen
- Nov 26, 2025FDA-FOIA-2025-13292Form 483Records released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 5 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.30Design controls1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.184Device history record1 of 1
- 21 CFR 820.198Complaint files1 of 1
- 21 CFR 801.45Devices that must be directly marked with a unique device identifier1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3014529181 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3014529181. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.