Prosec Protection Systems, Inc.
409 Joyce Kilmer Ave Ste 307, New Brunswick, New Jersey, United States
FEI 3003239512
FDA has 6 inspections on file for Prosec Protection Systems, Inc. in New Brunswick, New Jersey (FEI 3003239512), from 2011 to 2018. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 7 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Aug 2, 2013. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 6, 2011–2018
- 3 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 5 of 6
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers and contractors have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.
Documentation
Acceptance activities were not adequately documented.
Show 2 more citations
Acceptance status
The acceptance status of product was not maintained.
Personnel
Personnel do not have the necessary training to perform their jobs.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Show 2 more citations
Document review, approval by designated individual
Documents were not approved by designated individual(s) prior to issuance .
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Production processes
Production processes were not controlled to ensure that a device conforms to its specifications.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been established.
Show 1 more citation
Periodic equipment inspections
Periodic inspections of equipment conducted to ensure adherence to applicable maintenance schedules were not documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Lack of or inadequate organizational structure
The organizational structure has not been established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 820.25Personnel3 of 5
- 21 CFR 820.75Process validation3 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 of 5
- 21 CFR 820.20Management responsibility2 of 5
- 21 CFR 820.90Nonconforming product2 of 5
- 21 CFR 820.100Corrective and preventive action2 of 5
- 21 CFR 820.181Device master record2 of 5
- 21 CFR 820.22Quality audit1 of 5
Show all 14
- 21 CFR 820.40[Reserved]1 of 5
- 21 CFR 820.50Purchasing controls1 of 5
- 21 CFR 820.70Production and process controls1 of 5
- 21 CFR 820.86Acceptance status1 of 5
- 21 CFR 820.184Device history record1 of 5
- 21 CFR 820.198Complaint files1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3003239512 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3003239512. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.