Rite-dent Manufacturing Corp.

3750 E 10th Ct, Hialeah, Florida, United States

FEI 3003064146

FDA has 4 inspections on file for Rite-dent Manufacturing Corp. in Hialeah, Florida (FEI 3003064146), from 2010 to 2012. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Nov 16, 2010. FDA's compliance record also lists one other action under this FEI: Seizure, . No warning letter or import refusal is filed under this FEI.

Nine regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
4, 2010–2012
2 OAI · 0 VAI · 2 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design control - no procedures

      Procedures for design control have not been established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    3. Production processes

      Production processes were not developed, conducted, controlled, and monitored to ensure that a device conforms to its specifications.

    4. Equipment control activity documentation

      Equipment calibrations, inspections, checks, and maintenance activities have not been documented.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    Show 1 more citation
    1. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices23 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    2. Training records

      Personnel training is not documented.

    3. Design validation acceptance criteria

      Acceptance criteria were not established prior to the performance of validation activities.

    4. Management ensuring quality policy is understood

      Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.

    5. Quality plan

      A quality plan has not been adequately established.

    Show 18 more citations
    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    2. Design Validation - Risk analysis not performed/inadequate

      Risk analysis was not performed.

    3. Design verification - documentation

      The design verification results, including identification of the design, method(s), the date, and the individual(s) performing the verification, were not documented in the design history file.

    4. Personnel

      Personnel do not have the necessary training and experience to perform their jobs.

    5. Equipment control activity documentation

      Equipment calibrations and maintenance activities have not been documented.

    6. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    7. Sampling plans

      Sampling plans are not based on valid statistical rationale.

    8. Documentation

      In-process inspections, tests, or other verification activities and approvals were not documented.

    9. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    10. Lack of or inadequate DHR procedures

      Procedures for device history records have not been established.

    11. Lack of procedures, or not maintained

      Document control procedures have not been established and maintained.

    12. Documentation

      Acceptance activities were not documented and maintained as part of the device history record.

    13. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    14. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    15. Lack of or inadequate procedures for labeling

      Procedures to control labeling activities have not been adequately established.

    16. Lack of or inadequate organizational structure

      The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.

    17. Design control - no procedures

      Procedures for design control have not been established.

    18. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices14 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design control - no procedures

      Procedures to control the design process of the device were not established, documented, and implemented.

    2. Training

      Employees have not been adequately trained.

    3. Quality Audit/Reaudit - conducted

      Quality audits were not conducted at sufficient regular intervals, as prescribed by internal procedures to verify that the quality system is effective in fulfilling your quality system objectives.

    4. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not complete.

    5. DMR QA procedures and specifications

      The device master record does not include or refer to the location of quality assurance procedures and specifications.

    Show 9 more citations
    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance or rejection of incoming product were not implemented.

    2. DHR content

      The device history record does not demonstrate the device is manufactured in accordance with the device master record.

    3. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure that equipment is routinely calibrated and maintained were not established and documented.

    4. Design history file

      The design history file was not established or maintained.

    5. CAPA verification/validation of corrective/preventive action

      Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.

    6. DMR device specifications

      The device master record does not include or refer to the location of all specifications.

    7. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    8. Management responsibility

      Management with executive responsibility has not ensured that an adequate and effective quality system has been fully implemented and maintained at all levels of the organization.

    9. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by suppliers, contractors, and consultants were not established, defined, documented, and implemented.

Regulations FDA cited here

The 15 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

Show all 15

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3003064146 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3003064146. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.