Skip to content

DRG Instruments GmbH

Frauenbergstr. 18, Marburg, Germany

FEI 3002800697

FDA has 4 inspections on file for DRG Instruments GmbH in Marburg, Germany (FEI 3002800697), from 2010 to 2024. The latest ended on and was classified OAI, official action indicated. FDA has issued 2 warning letters under this FEI, the latest on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 4 citations from this inspection: 2 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 830, the unique device identification rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.

21 import refusals are filed under this FEI, the latest on . FDA’s Import Alert 89-08 Red List names this firm at this address. FDA printed no FEI on the entry. No court action is filed under this FEI.

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
4, 2010–2024
2 OAI · 1 VAI · 1 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Apr 29, 2025
Warning letters
· CDRH ↗
2 on file
Import refusals
21 refusals
Import alerts
89-08Red List
Matched on FDA’s firm name and street address

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Records of complaint investigation - UDI

      Records of complaint investigations do not include required information, including any unique device identifier (UDI) or universal product code (UPC).

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Devices subject to device identification GUDID data submission requirements.

      The information required by the Global Unique Device Identification Database for each medical device version or model required to bear a unique device identifier (UDI) was not provided.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been established.

    2. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.

    4. Report of violation of the Act (see 803.52(e)(9))

      A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.

    5. Remedial action

      When test/measurement equipment was found to not meet accuracy and precision limits, action was taken to evaluate whether there was any adverse effect on the device's quality.

    Show 1 more citation
    1. Document changes, review and approval, communication

      Changes to documents were not communicated to appropriate personnel in a timely manner.

  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    2. Product rework documentation, DHR {see also 820.184}

      Rework and reevaluation activities have not been documented in the device history record.

    3. Documentation

      Acceptance activities were not adequately documented and adequately maintained as part of the device history record.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Apr 29, 2025. It asked for a Form 483.

  • Inspection ending Nov 7, 2024OAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Apr 30, 2026. FDAgent holds no copy that FDA posted.

    1. Apr 29, 2025FDA-FOIA-2025-3808Form 483· 1 of 4 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 13 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

Show all 13

Warning letters and compliance actions

FDA Data Dashboard compliance actions

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 3002800697, one row per refused product line.

21 refusals from 2018-04-20 to 2025-07-22 · charged NO 510(K) on 16, NO PMA on 6 · no FDA sample analysis

FDA firm profile
  • 2025-07-22

    SYSTEM, TEST, RHEUMATOID FACTOR82DHR

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9350995-0/22/1

  • 2025-07-09

    RADIOIMMUNOASSAY, DEHYDROEPIANDROSTERONE (FREE AND SULFATE)75JKC

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9350953-9/181/1

  • 2025-07-09

    REAGENT, IMMUNOASSAY, IGG82KTO

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9350937-2/22/1

  • 2025-07-08

    ENZYME LINKED IMMUNOSORBENT ASSAY HERPES SIMPLEX VIRUS, HSV-283MYF

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9350980-2/92/1

  • 2025-07-08

    REAGENT, IMMUNOASSAY, IGG82KTO

    NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION

    Division of Northeast ImportsHTD-9350980-2/52/1

Show 16 more
  • 2025-07-08

    RADIOIMMUNOASSAY, ALDOSTERONE75CJM

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9350980-2/32/1

  • 2025-07-08

    ENZYME LINKED IMMUNOABSORBENT ASSAY, (CHLAMYDIAE GROUP)83LJC

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9350980-2/252/1

  • 2025-07-08

    RADIOIMMUNOASSAY, ANDROSTENEDIONE75CIZ

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9350980-2/242/1

  • 2025-07-08

    REAGENT, IMMUNOASSAY, IGG82KTO

    NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION

    Division of Northeast ImportsHTD-9350980-2/22/1

  • 2025-07-08

    REAGENT, IMMUNOASSAY, IGG82KTO

    NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION

    Division of Northeast ImportsHTD-9350980-2/212/1

  • 2025-07-08

    REAGENT, IMMUNOASSAY, IGG82KTO

    NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION

    Division of Northeast ImportsHTD-9350980-2/152/1

  • 2025-07-08

    RADIOIMMUNOASSAY, FREE THYROXINE75CEC

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9350980-2/122/1

  • 2025-07-08

    ASSAY, ENZYME LINKED IMMUNOSORBENT, PARVOVIRUS B19 IGG83MYL

    NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION

    Division of Northeast ImportsHTD-9350980-2/12/1

  • 2025-07-08

    ENZYME LINKED IMMUNOSORBENT ASSAY HERPES SIMPLEX VIRUS, HSV-283MYF

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9350980-2/102/1

  • 2025-07-03

    ENZYME LINKED IMMUNOABSORBENT ASSAY, CYTOMEGALOVIRUS83LFZ

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9351029-7/82/1

  • 2025-07-03

    ENZYME LINKED IMMUNOSORBENT ASSAY, HERPES SIMPLEX VIRUS, HSV-183MXJ

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9351029-7/62/1

  • 2025-07-03

    ENZYME LINKED IMMUNOSORBENT ASSAY HERPES SIMPLEX VIRUS, HSV-283MYF

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9351029-7/52/1

  • 2025-07-03

    ENZYME LINKED IMMUNOABSORBENT ASSAY, TOXOPLASMA GONDII83LGD

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9351029-7/122/1

  • 2025-05-08

    ENZYME IMMUNOASSAY, CORTISOL, SALIVARY75NHG

    NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9350866-3/141/1

  • 2019-04-16

    ENZYME IMMUNOASSAY, CORTISOL, SALIVARY75NHG

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9341625-5/121/1

  • 2018-04-20

    ENZYME IMMUNOASSAY, CORTISOL, SALIVARY75NHG

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Northeast ImportsHTD-9340418-6/81/1

Import alerts

FDA import alert pages, last synced Nov 23, 2026, 04:17 UTC. On a Red List, FDA can detain the firm's listed products without examining them.

  • 89-08Red ListDetention Without Physical Examination of Devices without Approved PMA's or IDE's and Other Devices Not Substantially Equivalent or Without a 510(k)Import alert pageAlert on fda.gov

    Date published 2017-10-25

    FDA’s entry
    DRG Instruments GmbH · Frauenbergstr. 18, Marburg, Hassia GERMANY
    FDA’s inspection record for FEI 3002800697
    DRG Instruments GmbH · Frauenbergstr. 18, Marburg, Germany

    Matched on exact firm name, street address and country. FDA printed no FEI on this entry.

    2 products

    • 75 N - - HGEnzyme Immunoassay, Cortisol, Salivary
    • 75 O - - TVResearch use only/clinical chemistry devices

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002800697 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002800697. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.