Skip to content

DDC Technologies, Inc.

311 Woods Ave, Oceanside, New York, United States

FEI 3002757272

FDA has 9 inspections on file for DDC Technologies, Inc. in Oceanside, New York (FEI 3002757272), from 2009 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 5 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the two FOIA requests for this site has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Nine regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
9, 2009–2024
0 OAI · 8 VAI · 1 NAI
Inspections with 483 observations shown
9 of 9
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Mar 4, 2025
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

9 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    4. Design history file

      The design history file was not established, does not demonstrate that the design was developed following the approved design plan and does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    5. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Failure to submit, distinct marking

      You did not submit a required product report prior to the introduction of the product into commerce.

    2. Incoming acceptance records, documentation

      Acceptance or rejection of incoming product was not documented.

    3. Failure to submit

      You did not submit an annual report for products requiring one.

    4. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Management review - Lack of or inadequate procedures

      Procedures for management review have not been adequately established.

    2. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    3. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been established.

    2. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    3. Quality Audit/Reaudit - conducted

      Quality audits have not been performed.

    4. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated and inspected have not been established.

    5. Failure to submit

      You did not submit an annual report for products requiring one.

    Show 3 more citations
    1. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.

    2. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    3. Design control - no procedures

      Procedures for design control have not been established.

  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    2. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    3. Design validation - documentation

      The results of design validation, including identification of the design, method(s), the date and the individual(s) performing validation, were not adequately documented in the design history file.

  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Complaints

      Complaint files are not maintained.

  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    2. Documented evaluation

      The evaluation of potential suppliers, contractors, and consultants was not documented.

    3. Complaints

      Complaint files are not maintained.

  8. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    2. Time to report - 10 days

      A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.

    3. Complaints

      Complaint files are not maintained.

    4. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

  9. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements were not established, defined, documented, complete, and implemented.

    2. Management review - Lack of or inadequate procedures

      Procedures for management review are not established, defined, documented, complete, and implemented.

    3. Procedures

      The procedures addressing documentation of corrective and preventive action activities were not established, defined, documented, complete, and implemented.

    4. Product rework procedures, Lack of or inadequate procedures

      Procedures for rework of nonconforming product were not established, defined, documented, complete, and implemented.

    5. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not established, defined, documented, and complete.

    Show 1 more citation
    1. Lack of Written MDR Procedures

      Written MDR procedures have not been developed, maintained, and implemented.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Mar 4, 2025. Each one asked for a Form 483.

  • Inspection ending Sep 6, 2024VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Mar 4, 2025FDA-FOIA-2025-1961Form 483· 1 of 5 sitesClosed
  • Inspection ending Oct 22, 2021NAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Jun 1, 2022FDA-FOIA-2022-4033Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 15 sections of 21 CFR in FDA’s published citations for this facility’s 9 inspections, with how many of them cited each.

Show all 15

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3002757272 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3002757272. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.