Philips Respironics, Inc.
1001 Murry Ridge Ln, Murrysville, Pennsylvania, United States
FEI 2518422
FDA has 16 inspections on file for Philips Respironics, Inc. in Murrysville, Pennsylvania (FEI 2518422), from 2010 to 2022. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the 10 FOIA requests for this site has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Nov 9, 2021. FDA has issued 3 warning letters under this FEI, the latest on (CDRH). FDA's compliance record also lists one other action under this FEI: Injunction, . 6 import refusals are filed under this FEI, the latest on .
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 16, 2010–2022
- 6 OAI · 3 VAI · 7 NAI
- Inspections with 483 observations shown
- 7 of 16
- Form 483s posted by FDA
- ↗
- Requested under FOIA
- 10 requests, latest FY2023
- Import refusals
- 6 refusals
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
16 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesPosted 483 · 8 citations
Read this Form 483 on its own page →Form 483 PDF on fda.gov
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Show 3 more citations
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Evaluation and Selection, Suppliers, Contractors, etc.
Potential consultants were not evaluated and selected based on their ability to meet specified requirements.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate receiving acceptance procedures
Procedures for acceptance of incoming product have not been established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Production processes
Production processes were not developed, conducted, controlled, and monitored to ensure that a device conforms to its specifications.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - simulated testing
The design was not validated under actual or simulated use conditions.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been adequately established.
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they are being used.
Show 7 more citations
Design Validation - Risk analysis not performed/inadequate
Risk analysis is incomplete.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
Lack of Written MDR Procedures
Written MDR procedures have not been maintained.
Lack of or inadequate procedures for labeling
Procedures to control labeling activities have not been adequately established.
Production processes
Production processes were not controlled to ensure that a device conforms to its specifications.
Quality System Procedures
Quality system procedures and instructions have not been established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
10 requests to FDA under the Freedom of Information Act named this site, most recently on FY2023.
Inspection ending May 19, 2022VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Nov 21, 2023FDA-FOIA-2023-10452Form 483No records found
Inspection ending Nov 9, 2021OAI
4 requests. In the inspection history above. FDA has posted its Form 483.
- Jul 25, 2022FDA-FOIA-2022-5389Form 483Withdrawn
- May 26, 2022FDA-FOIA-2022-3902EIRDenied
- Feb 11, 2022FDA-FOIA-2022-1134Form 483Closed
- Dec 1, 2021FDA-FOIA-2021-8242Form 483Released in part
Inspection ending Jan 10, 2020OAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Nov 21, 2023FDA-FOIA-2023-10451Form 483No records found
Inspection date not given
4 requests. FDA released records in answer to a request for its Form 483 on Jul 23, 2025. FDAgent holds no copy that FDA posted.
- FY2023FDA-FOIA-2023-6878Form 483· 1 of 3 sitesClosed
- Sep 30, 2025FDA-FOIA-2025-11488Form 483Open
- Jun 29, 2022FDA-FOIA-2022-4732Form 483Records released
- Jan 19, 2022FDA-FOIA-2022-455Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 16 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action4 of 7
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?3 of 7
- 21 CFR 820.30Design controls3 of 7
- 21 CFR 820.20Management responsibility2 of 7
- 21 CFR 820.40[Reserved]2 of 7
- 21 CFR 820.50Purchasing controls2 of 7
- 21 CFR 820.70Production and process controls2 of 7
- 21 CFR 820.198Complaint files2 of 7
Show all 14
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 7
- 21 CFR 806.10Reports of corrections and removals1 of 7
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 7
- 21 CFR 820.90Nonconforming product1 of 7
- 21 CFR 820.120Device labeling1 of 7
- 21 CFR 820.150Storage1 of 7
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- InjunctionCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2518422, one row per refused product line.
6 refusals from 2020-11-25 to 2026-05-04 · charged NOT LISTED on 5, REGISTERED on 4 · no FDA sample analysis
FDA firm profile- 2026-05-04
VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK
NOT LISTED 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southwest Imports799-8662407-6/60002/1
- 2026-05-04
VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK
NOT LISTED 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southwest Imports799-8662407-6/50002/1
- 2026-05-04
VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK
NOT LISTED 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southwest Imports799-8662407-6/40002/1
- 2026-05-04
VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK
NOT LISTED 502(o), 801(a)(3); MISBRANDING
REGISTERED 502(o), 801(a)(3); MISBRANDING
NO 510(K) 801(a)(3); 502(o) Misbranding
Division of Southwest Imports799-8662407-6/30002/1
- 2021-01-27
STANDARD POLYSOMNOGRAPH WITH ELECTROENCEPHALOGRAPH84OLV
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-9995762-2/11/1
Show 1 more
- 2020-11-25
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT73MNT
NOT LISTED 502(o), 801(a)(3); MISBRANDING
FRNMFGREG 502(o), 801(a)(3); MISBRANDING
Division of Southeast ImportsAEK-8092648-5/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 2518422 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 2518422. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.