Philips Respironics, Inc.

1001 Murry Ridge Ln, Murrysville, Pennsylvania, United States

FEI 2518422

FDA has 16 inspections on file for Philips Respironics, Inc. in Murrysville, Pennsylvania (FEI 2518422), from 2010 to 2022. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One of the 10 FOIA requests for this site has asked FDA for this inspection's 483.

The most recent OAI inspection ended on Nov 9, 2021. FDA has issued 3 warning letters under this FEI, the latest on (CDRH). FDA's compliance record also lists one other action under this FEI: Injunction, . 6 import refusals are filed under this FEI, the latest on .

Five regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
16, 2010–2022
6 OAI · 3 VAI · 7 NAI
Inspections with 483 observations shown
7 of 16
Form 483s posted by FDA
↗
Requested under FOIA
10 requests, latest FY2023
Warning letters
· CDRH ↗
3 on file
Import refusals
6 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

16 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesPosted 483 · 8 citations

    Read this Form 483 on its own page →Form 483 PDF on fda.gov

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Management ensuring quality policy is understood

      Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Design validation - user needs and intended uses

      Design validation did not ensure the device conforms to defined user needs and intended uses.

    4. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    5. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    Show 3 more citations
    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

    2. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential consultants were not evaluated and selected based on their ability to meet specified requirements.

    3. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation- Lack of or inadequate procedures

      Procedures for design validation have not been adequately established.

    2. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been established.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
  7. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  8. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  9. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Production processes

      Production processes were not developed, conducted, controlled, and monitored to ensure that a device conforms to its specifications.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

  10. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  11. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  12. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  13. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  14. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

  15. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design validation - simulated testing

      The design was not validated under actual or simulated use conditions.

    2. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    3. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    4. Lack of or inadequate procedures for storage.

      Procedures for the control of storage areas and stock rooms have not been adequately established.

    5. Not approved or obsolete document retrieval

      Documents that were not approved were observed at a location where they are being used.

    Show 7 more citations
    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Maintenance schedule, Lack of or inadequate schedule

      Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been maintained.

    5. Lack of or inadequate procedures for labeling

      Procedures to control labeling activities have not been adequately established.

    6. Production processes

      Production processes were not controlled to ensure that a device conforms to its specifications.

    7. Quality System Procedures

      Quality system procedures and instructions have not been established.

  16. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

10 requests to FDA under the Freedom of Information Act named this site, most recently on FY2023.

  • Inspection ending May 19, 2022VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Nov 21, 2023FDA-FOIA-2023-10452Form 483No records found
  • Inspection ending Nov 9, 2021OAI

    4 requests. In the inspection history above. FDA has posted its Form 483.

    1. Jul 25, 2022FDA-FOIA-2022-5389Form 483Withdrawn
    2. May 26, 2022FDA-FOIA-2022-3902EIRDenied
    3. Feb 11, 2022FDA-FOIA-2022-1134Form 483Closed
    4. Dec 1, 2021FDA-FOIA-2021-8242Form 483Released in part
  • Inspection ending Jan 10, 2020OAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Nov 21, 2023FDA-FOIA-2023-10451Form 483No records found
  • Inspection date not given

    4 requests. FDA released records in answer to a request for its Form 483 on Jul 23, 2025. FDAgent holds no copy that FDA posted.

    1. FY2023FDA-FOIA-2023-6878Form 483· 1 of 3 sitesClosed
    2. Sep 30, 2025FDA-FOIA-2025-11488Form 483Open
    3. Jun 29, 2022FDA-FOIA-2022-4732Form 483Records released
    4. Jan 19, 2022FDA-FOIA-2022-455Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for 7 of this facility’s 16 inspections, with how many of them cited each.

Show all 14

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • InjunctionCDRH · DevicesFDA record
  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 2518422, one row per refused product line.

6 refusals from 2020-11-25 to 2026-05-04 · charged NOT LISTED on 5, REGISTERED on 4 · no FDA sample analysis

FDA firm profile
  • 2026-05-04

    VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southwest Imports799-8662407-6/60002/1

  • 2026-05-04

    VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southwest Imports799-8662407-6/50002/1

  • 2026-05-04

    VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southwest Imports799-8662407-6/40002/1

  • 2026-05-04

    VENTILATOR, CONTINUOUS (RESPIRATOR)73CBK

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    NO 510(K) 801(a)(3); 502(o) Misbranding

    Division of Southwest Imports799-8662407-6/30002/1

  • 2021-01-27

    STANDARD POLYSOMNOGRAPH WITH ELECTROENCEPHALOGRAPH84OLV

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-9995762-2/11/1

Show 1 more
  • 2020-11-25

    VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT73MNT

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    FRNMFGREG 502(o), 801(a)(3); MISBRANDING

    Division of Southeast ImportsAEK-8092648-5/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2518422 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2518422. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.