PMT Corporation

1500 Park Rd, Chanhassen, Minnesota, United States

FEI 2182979

FDA has 5 inspections on file for PMT Corporation in Chanhassen, Minnesota (FEI 2182979), from 2009 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.198(c) and 21 CFR 820.72(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
5, 2009–2024
0 OAI · 3 VAI · 2 NAI
Inspections with 483 observations shown
4 of 5
Form 483s posted by FDA
None held for this FEI
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    2. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated have not been adequately established.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

  2. NAINo objectionable conditions that need follow-upDevices · Compliance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Production and Process Change Procedures, lack of or Inad.

      Procedures for changes to a procedure have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  3. NAINo objectionable conditions that need follow-upDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Complaints

      Complaint files are not adequately maintained.

    2. Production processes

      Production processes were not conducted and controlled to ensure that a device conforms to its specifications.

    3. Examination for accuracy

      Labeling was not sufficiently examined by a designated individual for accuracy including the correct expiration date and control numbers before release.

    4. DHR - not or inadequately maintained

      A device history record has not been adequately maintained.

    5. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    Show 2 more citations
    1. Equipment suitability & capability

      Certain measuring equipment is not suitable for its intended purposes.

    2. Records not kept

      There is no record maintained of a correction or removal action that was not required to be reported to FDA.

  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  5. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Analysis of data/reports from data sources

      Not all data from quality data sources are analyzed to identify existing and potential causes of nonconforming product and other quality problems.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.

    3. Documentation of disposition, justification, signature

      There is incomplete documentation of the justification for use of nonconforming product.

    4. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    5. Design verification procedures

      Procedures for verifying that design output meets design input were not complete.

    Show 1 more citation
    1. Lack of Written MDR Procedures

      Written MDR procedures have not been implemented.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.

Show all 12

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 2182979 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 2182979. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.