Horiba Instruments Incorporated

5449 Research Dr, Canton, Michigan, United States

FEI 1827821

FDA has 7 inspections on file for Horiba Instruments Incorporated in Canton, Michigan (FEI 1827821), from 2009 to 2017. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

The most recent OAI inspection ended on Jan 15, 2010. No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.30(c) and 21 CFR 820.75(a).

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
7, 2009–2017
1 OAI · 1 VAI · 5 NAI
Inspections with 483 observations shown
2 of 7
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  2. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  3. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
  5. NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
  6. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Calibration documentation

      There is no documentation of the equipment identification, the individual performing each calibration, and the next calibration date for inspection, measurement, and test equipment .

    2. Verification or validation of changes

      Changes to a process were not verified or validated.

    3. Design validation - production units

      The design was not validated using production units or their equivalents.

    4. Design validation - simulated testing

      The design was not validated using production units under actual or simulated use conditions.

    5. Review of data and documentation

      Finished devices were released for distribution before the associated data and documentation was reviewed.

    Show 4 more citations
    1. Techniques not used

      Statistical techniques are not used for control purposes where statistical techniques are applicable.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been validated and approved according to established procedures.

    3. Design input - documentation

      Design input requirements were not fully documented.

    4. DMR device specifications

      The device master record does not include or refer to the location of all specifications.

  7. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been fully validated and approved according to established procedures.

    2. Design history file

      The design history file was not established or maintained.

    3. Quality audit corrective action, reaudits {see also 820.100}

      Quality reaudits are not adequately performed.

    4. Investigation procedures

      The corrective and preventive action procedures addressing the investigation of the cause of nonconformities relating to product, processes, and the quality system were not complete.

    5. Evaluation and Selection, Suppliers, Contractors, etc.

      Potential contractors were not sufficiently evaluated and selected on the basis of their ability to meet specified requirements.

    Show 4 more citations
    1. Analysis of data/reports from data sources

      Not all data from quality data sources are analyzed to identify existing and potential causes of nonconforming product and other quality problems.

    2. Report of violation of the Act (see 803.52(e)(9))

      A violation of the FD&C Act involving a device which might present a risk to health was not reported to FDA.

    3. Design input procedures - appropriateness

      Procedures to ensure that a device's design input requirements are appropriate and address its intended use, including user/patient needs, were not complete.

    4. CAPA verification/validation of corrective/preventive action

      Corrective and preventive actions have not been verified or validated to ensure that the action is effective and does not adversely affect the finished device.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for 2 of this facility’s 7 inspections, with how many of them cited each.

Show all 11

Other FEIs recorded as Horiba Instruments Incorporated

Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1827821 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1827821. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.