FDA has 7 inspections on file for Biomet, Inc. in Warsaw, Indiana (FEI 1825034), from 2009 to 2023. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA has not published the citations from this inspection, and FDAgent holds no Form 483 that FDA posted for it. FDA posts only some 483s, so this does not show that none was issued. One of the two FOIA requests for this site has asked FDA for this inspection's 483.

FDA has issued 2 warning letters under this FEI, the latest on (CDRH). 6 import refusals are filed under this FEI, the latest on . No court action is filed under this FEI.

10 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
7, 2009–2023
0 OAI · 3 VAI · 4 NAI
Inspections with 483 observations shown
5 of 7
3 posted 483s not matched to an inspection
Form 483s posted by FDA
↗
3 posted
Requested under FOIA
2 requests, latest Jan 4, 2024
Warning letters
· CDRH ↗
2 on file
Import refusals
6 refusals
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

7 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC · FDA reading room, last synced Nov 23, 2026, 04:33 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices8 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design control - no procedures

      Procedures for design control have not been established.

    2. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    4. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    Show 3 more citations
    1. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    2. Lack of or inadequate In-process acceptance procedures

      Procedures for the acceptance of in-process product have not been adequately established.

    3. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

  3. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  4. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  5. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices11 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.

    2. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    3. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

    4. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    5. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    Show 6 more citations
    1. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    2. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    4. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been established.

    5. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    6. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

  6. Not classified yetFDA posted this Form 483; FDAgent has not matched it to a Data Dashboard inspectionNo inspection matchedForm 483 pagePDF
  7. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices14 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Environmental control Lack of or inadequate procedures

      Procedures to control environmental conditions have not been adequately established.

    2. Design control - no procedures

      Procedures for design control have not been established.

    3. Buildings

      Buildings are not of suitable design to perform necessary operations.

    4. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

    5. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

    Show 9 more citations
    1. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    4. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    6. Sampling plans

      Sampling plans are not based on valid statistical rationale.

    7. Process control procedures, Lack of or inadequate procedures

      Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

    8. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    9. Product rework procedures, Lack of or inadequate procedures

      Procedures for rework of nonconforming product have not been adequately established.

  8. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack/Inad procedure-Monitoring/Control of Validated Proces

      Procedures for monitoring and control of process parameters for a validated process have not been established.

    2. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

  9. NAINo objectionable conditions that need follow-upDevices · Bioresearch MonitoringNone
  10. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is incomplete.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Jan 4, 2024. Each one asked for a Form 483.

  • Inspection ending Nov 9, 2023VAI

    1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Jan 4, 2024FDA-FOIA-2024-133Form 483Closed
  • Inspection ending Jul 30, 2020

    1 request. Not listed in FDA’s Data Dashboard yet. None of FDA’s log entries for these requests say records were released.

    1. Nov 28, 2023FDA-FOIA-2023-10625Form 483No records found

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 12 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 7 inspections, with how many of them cited each.

Show all 12

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched some of these, so the link opens FDA’s record of the action.

Import refusals

FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1825034, one row per refused product line.

6 refusals from 2020-01-21 to 2020-04-17 · charged REGISTERED on 5, NOT LISTED on 1 · no FDA sample analysis

FDA firm profile
  • 2020-04-17

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-6484848-3/31/5

  • 2020-04-17

    PLATE, FIXATION, BONE87HRS

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-6484848-3/31/3

  • 2020-04-17

    PLATE, BONE76JEY

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-6484848-3/31/1

  • 2020-01-21

    SCREW, FIXATION, BONE87HWC

    NOT LISTED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-4888555-1/11/1

  • 2020-01-21

    SCREW, FIXATION, BONE87HWC

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-4706784-7/21/1

Show 1 more
  • 2020-01-21

    SCREW, FIXATION, BONE87HWC

    REGISTERED 502(o), 801(a)(3); MISBRANDING

    Division of Southeast Imports799-4706784-7/11/1

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1825034 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1825034. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.