Zimmer Biomet, Inc. Form 483, April 2018
FDA issued this Form 483 to Zimmer Biomet, Inc. in Warsaw, Indiana (FEI 1825034) at the end of a device inspection that ended April 24, 2018. FDA classified the inspection No Action Indicated (NAI).
A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.
Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.
- Inspection ended
- Apr 24, 2018
- Devices · Postmarket Assurance: Devices
- FDA's outcome
- NAI
- No Action Indicated
- Posted by FDA
- Jan 17, 2024
- 2,094 days after the inspection
FDA's 11 citations from this inspection
FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
After this inspection
FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.
- Warning LetterAug 24, 2018CDRH · 122 days after this inspection ended
Other Form 483s FDA posted for this facility
Every Form 483 FDA posts under FEI 1825034, newest first.
- Form 483, May 2018FDA record date May 29, 2018 · Posted Jan 17, 2024
- Form 483, November 2016Inspection ended Nov 22, 2016 · Posted Jan 17, 2024NAI
Ask FDAgent about this Form 483
FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.
Related
About this data
- The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
- FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.
- FDAgent leaves out the names of the people the Form 483 is addressed to and of the investigators.