Skip to content

Zimmer Biomet, Inc. Form 483, April 2018

FDA issued this Form 483 to Zimmer Biomet, Inc. in Warsaw, Indiana (FEI 1825034) at the end of a device inspection that ended April 24, 2018. FDA classified the inspection No Action Indicated (NAI).

A Form 483 lists the investigators' observations at the end of an inspection. It is not FDA's final determination.

Source: FDA's Electronic Reading Room, posted Jan 17, 2024. Form 483 PDF on fda.gov. FDAgent is independent of FDA.

Inspection ended
Apr 24, 2018
Devices · Postmarket Assurance: Devices
FDA's outcome
NAI
No Action Indicated
Posted by FDA
Jan 17, 2024
2,094 days after the inspection
Facility
FEI 1825034
Warsaw, Indiana

FDA's 11 citations from this inspection

FDA's citation titles and text from its Data Dashboard. These are FDA's summaries, not the Form 483 itself. FDAgent hasn't verified a reading of this PDF, so the Form 483's own words are in the PDF only.

  1. Calibration, Inspection, etc. Procedures Lack of or Inadequ

    Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.

  2. Process control procedures, Lack of or inadequate procedures

    Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been adequately established.

  3. Lack/Inad procedure-Monitoring/Control of Validated Proces

    Procedures for monitoring and control of process parameters for a validated process have not been adequately established.

  4. Lack of or inadequate procedures

    Procedures for corrective and preventive action have not been adequately established.

  5. Lack of or inadequate process validation

    A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

  6. Purchasing controls, Lack of or inadequate procedures

    Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

  7. Lack of or inadequate procedures - Acceptance activities

    Procedures for acceptance activities have not been adequately established.

  8. Nonconforming product, Lack of or inadequate procedures

    Procedures have not been established to control product that does not conform to specified requirements.

  9. Training - Lack of or inadequate procedures

    Procedures for training and identifying training needs have not been established.

  10. Environmental control Lack of or inadequate procedures

    Procedures to control environmental conditions have not been adequately established.

  11. Design Validation - Risk analysis not performed/inadequate

    Risk analysis is inadequate.

After this inspection

FDA's compliance actions for this FEI dated after this inspection and before its next one. FDAgent links a letter when FDA's case number ties it to this facility.

  • Warning LetterAug 24, 2018CDRH · 122 days after this inspection ended

Other Form 483s FDA posted for this facility

Every Form 483 FDA posts under FEI 1825034, newest first.

All inspections at this facility →

Ask FDAgent about this Form 483

FDAgent reads the 483, the facility's inspections, warning letters and FDA's citations and answers in plain English, with links to the records.

Edit it or ask as written. 3 questions free, no card needed.

About this data

  • The PDF and its dates come from FDA's Electronic Reading Room, last synced Nov 23, 2026, 04:33 UTC; the inspection, its outcome and FDA's citations from FDA's Data Dashboard.
  • FDAgent shows a Form 483’s own text only when its reading of that PDF passes every check.