Mesa Laboratories, Inc.
12100 W 6th Ave, Lakewood, Colorado, United States
FEI 1720309
FDA has 3 inspections on file for Mesa Laboratories, Inc. in Lakewood, Colorado (FEI 1720309), from 2012 to 2018. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
1 import refusal is filed under this FEI, on . No warning letter or court action is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.198(a), 21 CFR 820.70(i) and 21 CFR 820.90(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 3, 2012–2018
- 0 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Servicing - Lack of or inadequate procedures
Procedures or instructions for performing servicing activities and verifying that servicing meets specified requirements have not been adequately established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Software validation for automated processes
Software used as part of production has not been validated for its intended use according to an established protocol.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Not approved or obsolete document retrieval
Documents that were not approved were observed at a location where they are being used.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design validation - software validation not performed
Validation of device software was not performed.
Show 7 more citations
Documentation of software validation
Software validation activities and results for computers or automated data processing systems used as part of the quality system have not been documented.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been adequately established.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been adequately established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Complaints
Complaint files are not adequately maintained.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.90Nonconforming product3 of 3
- 21 CFR 820.198Complaint files3 of 3
- 21 CFR 820.70Production and process controls2 of 3
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.30Design controls1 of 3
- 21 CFR 820.40[Reserved]1 of 3
- 21 CFR 820.50Purchasing controls1 of 3
- 21 CFR 820.75Process validation1 of 3
Show all 13
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
- 21 CFR 820.181Device master record1 of 3
- 21 CFR 820.184Device history record1 of 3
- 21 CFR 820.200Servicing1 of 3
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1720309, one row per refused product line.
1 refusal on 2024-04-11 · charged NOT LISTED on 1 · no FDA sample analysis
FDA firm profile- 2024-04-11
STERILIZER, DRY HEAT80KMH
NOT LISTED 502(o), 801(a)(3); MISBRANDING
Division of Southeast Imports799-9970097-4/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1720309 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1720309. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.