FDA has 5 inspections on file for Ohio Medical Corporation in Gurnee, Illinois (FEI 1419185), from 2009 to 2015. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 11 citations from this inspection: 10 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 803, the device adverse event reporting rules. They are listed below.
The most recent OAI inspection ended on May 19, 2009. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Nine regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2009–2015
- 1 OAI · 3 VAI · 1 NAI
- Inspections with 483 observations shown
- 5 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Process validation activities have not been documented.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
DMR - not or inadequately maintained
A device master record has not been maintained.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 6 more citations
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Supplier notification of changes
There is no agreement with suppliers to notify you of changes in the product or service.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of procedures, or not maintained
Document control procedures have not been established and maintained.
Evaluation, timeliness, identification
Complaints representing events that are MDR reportable were not promptly reviewed, evaluated, and investigated by a designated individual.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Show 1 more citation
Reporting Information Reasonably Known
An MDR report submitted to FDA did not include all information that was reasonably known to the manufacturer.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Procedures for product review,disposition lack of/inadequate
Procedures that define the responsibility for review and the authority for the disposition of nonconforming product have not been adequately established.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals and at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
Show 4 more citations
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Management review - defined interval, sufficient frequency
Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR content
The device history record does not demonstrate the device is manufactured in accordance with the device master record.
Uniform and timely processing
Complaint handling procedures have not been implemented to ensure that all complaints are processed in a uniform and timely manner.
Nonconforming product evaluation/investigation
The evaluation of nonconforming product has not been documented.
Documentation
Acceptance activities were not fully documented.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Show 1 more citation
General
Complaint handling procedures for receiving complaints have not been defined.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action5 of 5
- 21 CFR 820.198Complaint files5 of 5
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?3 of 5
- 21 CFR 820.30Design controls3 of 5
- 21 CFR 820.90Nonconforming product3 of 5
- 21 CFR 820.22Quality audit2 of 5
- 21 CFR 820.184Device history record2 of 5
- 21 CFR 820.20Management responsibility1 of 5
Show all 13
- 21 CFR 820.40[Reserved]1 of 5
- 21 CFR 820.50Purchasing controls1 of 5
- 21 CFR 820.75Process validation1 of 5
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 5
- 21 CFR 820.181Device master record1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1419185 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1419185. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.