Richard Wolf Medical Instruments Corp.
353 Corporate Woods Pkwy, Vernon Hills, Illinois, United States
FEI 1418479
FDA has 9 inspections on file for Richard Wolf Medical Instruments Corp. in Vernon Hills, Illinois (FEI 1418479), from 2009 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
The most recent OAI inspection ended on May 8, 2013. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 9, 2009–2025
- 1 OAI · 4 VAI · 4 NAI
- Inspections with 483 observations shown
- 6 of 9
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
9 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Document change records, maintained.
Records of changes to documents were not adequately maintained.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Equipment control activity documentation
Equipment maintenance activities have not been documented.
Show 1 more citation
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Records of MDR Investigation
Investigation records of MDR reportable complaints do not include required information.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Show 3 more citations
Conducting an Investigation
An MDR report with incomplete information was submitted to FDA without a statement explaining why such information was incomplete and the steps taken to obtain the information.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Acceptable supplier records, inadequate records
Records of acceptable suppliers have not been adequately established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Supplier oversight
The type and extent of control to be exercised over the product, services, suppliers, contractors, and consultants was not clearly defined.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers, contractors, and consultants have not been adequately established.
Report of Death or Serious Injury
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
DHR - not or inadequately maintained
A device history record has not been maintained.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated and approved according to established procedures.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities were not complete.
Process control procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications were not implemented.
Identification of actions needed
Not all of the actions needed to correct and prevent the recurrence of nonconforming product and other quality problems have been identified.
Show 5 more citations
Calibration documentation
There is no documentation of calibration dates for measurement equipment .
ID label, labeling
The device history record does not include the primary identification label and labeling for each device.
Investigation
Certain indicators of nonconformities are not investigated to determine the cause of the nonconformity.
Documentation
Corrective and preventive action activities have not been documented, including investigations of causes of nonconformities.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed and investigated where necessary.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for 6 of this facility’s 9 inspections, with how many of them cited each.
- 21 CFR 820.198Complaint files4 of 6
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?3 of 6
- 21 CFR 820.50Purchasing controls3 of 6
- 21 CFR 820.100Corrective and preventive action3 of 6
- 21 CFR 820.20Management responsibility2 of 6
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 6
- 21 CFR 820.184Device history record2 of 6
- 21 CFR 820.22Quality audit1 of 6
Show all 14
- 21 CFR 820.25Personnel1 of 6
- 21 CFR 820.40[Reserved]1 of 6
- 21 CFR 820.70Production and process controls1 of 6
- 21 CFR 820.75Process validation1 of 6
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 6
- 21 CFR 820.90Nonconforming product1 of 6
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1418479 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1418479. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.