EyeKon Medical, Inc.
2451 Enterprise Rd, Clearwater, Florida, United States
FEI 1038833
FDA has 5 inspections on file for EyeKon Medical, Inc. in Clearwater, Florida (FEI 1038833), from 2008 to 2025. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 4 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Oct 3, 2008. FDA issued a warning letter on (CDRH). 1 import refusal is filed under this FEI, on . No court action is filed under this FEI.
Three regulations were cited at more than one inspection: 21 CFR 820.100(a), 21 CFR 820.198(a) and 21 CFR 820.75(a).
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2008–2025
- 1 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 3 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Aug 12, 2025
- Warning letters
- · CDRH ↗
- Import refusals
- 1 refusal
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Supplier oversight
The type and extent of control to be exercised over consultants was not clearly defined.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 2 more citations
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices9 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been validated according to established procedures.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Production and Process Change Procedures, lack of or Inad.
Procedures for changes to a specification, method, process, and procedure have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Acceptance status
The acceptance status of product was not identified to indicate conformance or nonconformance with acceptance criteria.
Show 4 more citations
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Design history file
The design history file was not established.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: DevicesCitations not published
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Aug 12, 2025. It asked for a Form 483.
Inspection ending Jun 10, 2025NAI
1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 16, 2025. FDAgent holds no copy that FDA posted.
- Aug 12, 2025FDA-FOIA-2025-9630Form 483· 1 of 4 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 12 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 3
- 21 CFR 820.198Complaint files3 of 3
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.50Purchasing controls2 of 3
- 21 CFR 820.70Production and process controls2 of 3
- 21 CFR 820.75Process validation2 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.22Quality audit1 of 3
Show all 12
- 21 CFR 820.25Personnel1 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 3
- 21 CFR 820.86Acceptance status1 of 3
- 21 CFR 820.181Device master record1 of 3
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Import refusals
FDA Data Dashboard, last synced Nov 23, 2026, 04:17 UTC. Shipments FDA refused entry into the US, filed under FEI 1038833, one row per refused product line.
1 refusal on 2022-08-05 · charged NOT LISTED on 1, NO PMA on 1 · no FDA sample analysis
FDA firm profile- 2022-08-05
LENS, INTRAOCULAR86HQL
NOT LISTED 502(o), 801(a)(3); MISBRANDING
NO PMA 501(f)(1)(B), 801(a)(3); ADULTERATION
Division of Southeast ImportsAEK-1580052-5/11/1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 1038833 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 1038833. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.