Unimed Surgical Products, Inc.

10401 Belcher Rd S, Largo, Florida, United States

FEI 1000307745

FDA has 3 inspections on file for Unimed Surgical Products, Inc. in Largo, Florida (FEI 1000307745), from 2009 to 2013. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 10 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.

The most recent OAI inspection ended on Sep 21, 2011. No warning letter, court action or import refusal is filed under this FEI.

10 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
3, 2009–2013
2 OAI · 0 VAI · 1 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely checked and maintained have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    4. Lack of or inadequate process validation

      A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.

    5. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    Show 5 more citations
    1. Procedures not adequately established or maintained

      Document control procedures have not been adequately maintained.

    2. Design control - no procedures

      Procedures for design control have not been established.

    3. Quality audits - auditor independence

      Individuals who conduct quality audits have direct responsibility for the matters being audited.

    4. Investigation of device failures

      Complaints involving the possible failure of a device and packaging to meet any of its specifications were not evaluated where necessary.

    5. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

  2. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    2. Calibration procedures - content

      Calibration procedures do not include specific directions and limits for accuracy and precision and provisions for remedial action.

    3. Training records

      Personnel training is not documented.

    4. Quality audits - auditor independence

      Individuals who conduct quality audits have direct responsibility for the matters being audited.

    5. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    Show 7 more citations
    1. Documentation

      Process validation have not been adequately documented and adequately approved.

    2. Document review, approval by designated individual

      Documents were not reviewed and not approved by designated individual(s) prior to issuance .

    3. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    5. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    6. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.

    7. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.

  3. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices13 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure that equipment is routinely calibrated were not complete and implemented.

    2. Process control instructions

      Process controls are not conducted in accordance with documented instructions and standard operating procedures.

    3. Design verification procedures

      Procedures for verifying that design output meets design input were not complete and implemented.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed, maintained, and implemented.

    5. Lack of or inadequate procedures

      The procedures for implementing corrective and preventive actions were not complete.

    Show 8 more citations
    1. Report of Death or Serious Injury

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device may have caused or contributed to a death or serious injury.

    2. Evaluation of suppliers, contractors, etc., requirements

      Adequate quality requirements that must be met by suppliers were not established, documented, and implemented.

    3. Investigation of device failures

      Complaints involving the possible failure of packaging to meet any of its specifications were not evaluated where necessary.

    4. Process control procedures, Lack of or inadequate procedures

      Process controls do not provide for documented instructions, standard operating procedures (SOPs), and methods that define and control the manner of production.

    5. Training records

      Employee training is not fully documented.

    6. Incomplete, ambiguous, conflicting design requirements

      Design input requirements that are conflicting were not addressed.

    7. Quality audits - Lack of or inadequate procedures

      Procedures for conducting quality audits were not complete.

    8. Documentation

      Process validation activities and results have not been documented.

Regulations FDA cited here

The 14 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Show all 14

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000307745 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000307745. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.