Oakworks Inc

923 E Wellspring Rd, New Freedom, Pennsylvania, United States

FEI 1000121370

FDA has 3 inspections on file for Oakworks Inc in New Freedom, Pennsylvania (FEI 1000121370), from 2014 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 6 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Seven regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
3, 2014–2024
0 OAI · 3 VAI · 0 NAI
Inspections with 483 observations shown
3 of 3
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Jul 30, 2024
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Radiation Control and Health Safety Act6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Design validation - simulated testing

      The design was not validated under actual or simulated use conditions.

    3. Complaints

      Complaint files are not adequately maintained.

    4. Quality audits - auditor independence

      Individuals who conduct quality audits have direct responsibility for the matters being audited.

    5. Design verification - documentation

      The design verification results, including identification of the design and method(s), were not adequately documented in the design history file.

    Show 1 more citation
    1. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system at defined intervals.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    2. Report of risk to health

      A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.

    3. Management review - defined interval, sufficient frequency

      Management with executive responsibility has not reviewed the suitability and effectiveness of the quality system with sufficient frequency.

    4. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

    5. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    Show 5 more citations
    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    2. Sampling plans

      Sampling plans are not based on valid statistical rationale.

    3. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    4. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record and 21 CFR 820.

    5. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices15 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    2. Lack of or inadequate receiving acceptance procedures

      Procedures for acceptance of incoming product have not been adequately established.

    3. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    4. Training - Lack of or inadequate procedures

      Procedures for training and identifying training needs have not been adequately established.

    5. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    Show 10 more citations
    1. Document review, approval documentation

      The documentation of approval of documents does not include the document approval date and the signature of the approving official.

    2. Documentation

      Acceptance activities were not adequately documented.

    3. Lack of or inadequate procedures for labeling

      Procedures to control labeling activities have not been established.

    4. Design plans - Lack of or inadequate

      Design plans that describe or reference the design and development activities and define responsibility for implementation have not been established.

    5. Design control - no procedures

      Procedures for design control have not been established.

    6. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    7. Sampling plans

      Sampling plans are not written and based on valid statistical rationale.

    8. Design input - documentation

      Design input requirements were not adequately documented.

    9. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    10. Design verification - output does not meet input requirement

      Design verification does not confirm that design output meets design input requirements.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Jul 30, 2024. Each one asked for a Form 483.

  • Inspection ending Jan 17, 2024VAI

    2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Jul 30, 2024FDA-FOIA-2024-6622Form 483· 1 of 2 sitesClosed
    2. Feb 22, 2024FDA-FOIA-2024-1950Form 483Closed

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 15 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.

Show all 15

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 1000121370 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 1000121370. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.