GPS DENTAL LAB, INC
3501 Parkway Center Ct, Orlando, Florida, United States
FEI 3020064003
FDA has one inspection on file for GPS DENTAL LAB, INC in Orlando, Florida (FEI 3020064003). It ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 12 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Three FOIA requests have asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 1, in 2024
- Inspections with 483 observations shown
- 1 of 1
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 3 requests, latest Sep 29, 2025
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
1 inspection · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Identification procedures, Lack of or inadequate procedures
Procedures for identifying product during all stages of receipt, production, distribution, and installation have not been adequately established.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Show 7 more citations
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked and maintained have not been adequately established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been established.
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
Three requests to FDA under the Freedom of Information Act named this site, most recently on Sep 29, 2025. Each one asked for a Form 483.
Inspection ending Nov 5, 2024VAI
3 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Jul 31, 2025. FDAgent holds no copy that FDA posted.
- Sep 29, 2025FDA-FOIA-2025-11417Form 483· 1 of 4 sitesRecords released
- Apr 30, 2025FDA-FOIA-2025-3889Form 483· 1 of 4 sitesRecords released
- Feb 20, 2025FDA-FOIA-2025-1593Form 483· 1 of 3 sitesRecords released
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for this facility’s 1 inspection.
- 21 CFR 820.20Management responsibility1 of 1
- 21 CFR 820.22Quality audit1 of 1
- 21 CFR 820.40[Reserved]1 of 1
- 21 CFR 820.50Purchasing controls1 of 1
- 21 CFR 820.60Identification1 of 1
- 21 CFR 820.70Production and process controls1 of 1
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 1
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 1
Show all 11
- 21 CFR 820.90Nonconforming product1 of 1
- 21 CFR 820.100Corrective and preventive action1 of 1
- 21 CFR 820.198Complaint files1 of 1
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3020064003 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3020064003. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.