Natus Medical Incorporated
5900 1st Ave S, Seattle, Washington, United States
FEI 3018859
FDA has 9 inspections on file for Natus Medical Incorporated in Seattle, Washington (FEI 3018859), from 2008 to 2018. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published one citation from this inspection, under 21 CFR Part 803, the device adverse event reporting rules. It is listed below.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Six regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 9, 2008–2018
- 0 OAI · 2 VAI · 7 NAI
- Inspections with 483 observations shown
- 5 of 9
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
9 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: Devices1 citation published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been adequately established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Show 7 more citations
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been adequately established.
Design transfer - Lack of or inadequate procedures
Procedures for design transfer have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
Records of complaint investigation
Records of complaint investigations do not include required information.
Design verification - Lack of or inadequate procedures
Procedures for design verification have not been adequately established.
Design input - documentation
Design input requirements were not adequately documented.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
DMR - not or inadequately maintained
A device master record has not been maintained.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Design history file
The design history file was not established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Show 2 more citations
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Design validation- Lack of or inadequate procedures
Procedures for design validation have not been established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design validation - simulated testing
The design was not validated under actual or simulated use conditions.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- NAINo objectionable conditions that need follow-upDevices · Project Evaluation: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Acceptance records
The device history record does not include complete acceptance records that demonstrate the device is manufactured in accordance with the device master record.
Calibration documentation
There is incomplete documentation of calibration dates for inspection, measurement, and test equipment .
Nonconforming product, Lack of or inadequate procedures
Procedures were not followed for the control of products that do not conform to specifications.
Design verification - documentation
The design verification results, including identification of the design, were not documented and filed in the design history file.
Product rework documentation, DHR {see also 820.184}
Rework and reevaluation activities have not been fully documented in the device history record.
Regulations FDA cited here
The 10 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 9 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls4 of 5
- 21 CFR 820.100Corrective and preventive action3 of 5
- 21 CFR 820.50Purchasing controls2 of 5
- 21 CFR 820.90Nonconforming product2 of 5
- 21 CFR 820.184Device history record2 of 5
- 21 CFR 820.198Complaint files2 of 5
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 5
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 5
Show all 10
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 5
- 21 CFR 820.181Device master record1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Other FEIs recorded as Natus Medical Incorporated
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Natus Medical IncorporatedFEI 3013242889Pleasanton, California, United States
Last inspected 2018-12-03NAI
Natus Medical IncorporatedFEI 2023988San Diego, California, United States
Last inspected 2016-04-29NAI
Natus Medical IncorporatedFEI 1450453Schaumburg, Illinois, United States
Last inspected 2015-03-30NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3018859 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3018859. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.