OndaMed, Inc.

5 Fairlawn Dr Ste 200, Washingtonville, New York, United States

FEI 3011088301

FDA has 4 inspections on file for OndaMed, Inc. in Washingtonville, New York (FEI 3011088301), from 2015 to 2026. The latest ended on and was classified NAI, no action indicated.

NAI means FDA found no objectionable conditions or practices, or found none that justified further action.

FDA published 2 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.

No warning letter, court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.72(a) and 21 CFR 820.184.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
NAI
Inspections on file
4, 2015–2026
0 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
4 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
1 request, latest Oct 6, 2026
Warning letters
None on file
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 8.2.6

      Evidence of conformity to the acceptance criteria not maintained

      Evidence of conformity to the acceptance criteria for product monitoring and measurement was not maintained.

    2. ISO 13485:2016 Clause 8.3.4

      Rework records not maintained

      Records of rework were not adequately maintained.

  2. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    2. Equipment control activity documentation

      Equipment calibrations have not been documented.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Reports and information documentation

      The written MDR procedure does not include documentation and recordkeeping requirements for all Medical Device Reports and information submitted to FDA.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design verification - documentation

      The design verification results, including identification of the design, method(s), the date and the individual(s) performing the verification, were not documented in the design history file.

    2. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been implemented.

    5. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated and maintained have not been adequately established.

  4. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices10 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Rationale documented for no investigation

      Records for complaints where no investigation was made do not include required information.

    2. Design validation - software validation not performed

      Validation of device software is incomplete.

    3. Design history file

      The design history file was not established, does not demonstrate that the design was developed following the approved design plan, and does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    4. DHR content

      The device history record does not demonstrate that the device was manufactured in accordance with the device master record.

    5. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been adequately established.

    Show 5 more citations
    1. Service reports

      Service reports do not include the required information.

    2. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    4. Supplier notification of changes

      There is no agreement with suppliers to notify you of changes in the product or service.

    5. Evaluation of suppliers, contractors, etc., requirements

      Requirements that must be met by suppliers have not been established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

One request to FDA under the Freedom of Information Act named this site, on Oct 6, 2026. It asked for a Form 483.

  • Inspection ending Jul 15, 2026NAI

    1 request. In the inspection history above.

    1. Oct 6, 2026FDA-FOIA-2026-9718Form 483· 1 of 5 sitesOpen

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 11 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.

Show all 11

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3011088301 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3011088301. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.