OndaMed, Inc.
5 Fairlawn Dr Ste 200, Washingtonville, New York, United States
FEI 3011088301
FDA has 4 inspections on file for OndaMed, Inc. in Washingtonville, New York (FEI 3011088301), from 2015 to 2026. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 2 citations from this inspection. They are listed below. One FOIA request has asked FDA for this inspection's 483.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.72(a) and 21 CFR 820.184.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 4, 2015–2026
- 0 OAI · 2 VAI · 2 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Oct 6, 2026
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
- ISO 13485:2016 Clause 8.2.6
Evidence of conformity to the acceptance criteria not maintained
Evidence of conformity to the acceptance criteria for product monitoring and measurement was not maintained.
- ISO 13485:2016 Clause 8.3.4
Rework records not maintained
Records of rework were not adequately maintained.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Equipment control activity documentation
Equipment calibrations have not been documented.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Reports and information documentation
The written MDR procedure does not include documentation and recordkeeping requirements for all Medical Device Reports and information submitted to FDA.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices5 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - documentation
The design verification results, including identification of the design, method(s), the date and the individual(s) performing the verification, were not documented in the design history file.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been implemented.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated and maintained have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices10 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Rationale documented for no investigation
Records for complaints where no investigation was made do not include required information.
Design validation - software validation not performed
Validation of device software is incomplete.
Design history file
The design history file was not established, does not demonstrate that the design was developed following the approved design plan, and does not demonstrate that the design was developed following the requirements of 21 CFR 820.
DHR content
The device history record does not demonstrate that the device was manufactured in accordance with the device master record.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Show 5 more citations
Service reports
Service reports do not include the required information.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been adequately established to control product that does not conform to specified requirements.
Supplier notification of changes
There is no agreement with suppliers to notify you of changes in the product or service.
Evaluation of suppliers, contractors, etc., requirements
Requirements that must be met by suppliers have not been established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Oct 6, 2026. It asked for a Form 483.
Inspection ending Jul 15, 2026NAI
1 request. In the inspection history above.
- Oct 6, 2026FDA-FOIA-2026-9718Form 483· 1 of 5 sitesOpen
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 4
- 21 CFR 820.30Design controls2 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 4
- 21 CFR 820.184Device history record2 of 4
- 21 CFR 820.198Complaint files2 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance1 of 4
- 21 CFR 820.90Nonconforming product1 of 4
Show all 11
- 21 CFR 820.100Corrective and preventive action1 of 4
- 21 CFR 820.181Device master record1 of 4
- 21 CFR 820.200Servicing1 of 4
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3011088301 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3011088301. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.