Skip to content

US Vascular, Inc.

7874 Sw Nimbus Ave, Beaverton, Oregon, United States

FEI 3010089756

FDA has 6 inspections on file for US Vascular, Inc. in Beaverton, Oregon (FEI 3010089756), from 2016 to 2024. The latest ended on and was classified VAI, voluntary action indicated.

VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.

FDA published 3 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. Two of the three FOIA requests for this site have asked FDA for this inspection's 483.

The most recent OAI inspection ended on Mar 30, 2018. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.

12 regulations were cited at more than one inspection; they are listed below.

Sources:FDA firm profile ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
VAI
Inspections on file
6, 2016–2024
2 OAI · 2 VAI · 2 NAI
Inspections with 483 observations shown
6 of 6
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
3 requests, latest Apr 14, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    2. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices1 citation published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

  3. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices3 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    2. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    3. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

  4. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of procedures, or not maintained

      Document control procedures have not been established.

    2. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    3. Lack of Written MDR Procedures

      Written MDR procedures have not been implemented.

    4. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been established.

    5. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    Show 7 more citations
    1. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been established.

    2. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been established.

    3. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    4. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    5. DMR - not or inadequately maintained

      A device master record has not been maintained.

    6. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

    7. Lack of or inadequate DHR procedures

      Procedures for device history records have not been established.

  5. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices14 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Lack of or inadequate procedures - Acceptance activities

      Procedures for acceptance activities have not been adequately established.

    2. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.

    3. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    4. Lack of System for Timely Submission of Reports

      The written MDR procedure does not include an internal system which provides for timely transmission of complete medical device reports to FDA.

    5. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been adequately established.

    Show 9 more citations
    1. Design history file

      The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.

    2. Calibration procedures - content

      Calibration procedures do not include specific directions and limits for accuracy and precision.

    3. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    4. Lack of or inadequate DHR procedures

      Procedures for device history records have not been adequately established.

    5. Procedures not adequately established or maintained

      Document control procedures have not been adequately established.

    6. Lack of System for Determining MDR Events

      The written MDR procedure does not include an internal system which provides for a standardized review process/procedure for determining when an event meets the criteria for reporting.

    7. Design changes - Lack of or Inadequate Procedures

      Procedures for design change have not been adequately established.

    8. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.

    9. DMR - not or inadequately maintained

      A device master record has not been adequately maintained.

  6. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices12 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design control - no procedures

      Procedures for design control have not been established.

    2. Calibration, Inspection, etc. Procedures Lack of or Inadequ

      Procedures to ensure equipment is routinely calibrated and maintained have not been established.

    3. Quality policy and objectives

      The quality policy and quality objectives was/were not established by management with executive responsibility.

    4. Lack of Written MDR Procedures

      Written MDR procedures have not been developed.

    5. Purchasing controls, Lack of or inadequate procedures

      Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.

    Show 7 more citations
    1. Quality audits - Lack of or inadequate procedures

      Procedures for quality audits have not been established.

    2. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been established to control product that does not conform to specified requirements.

    3. Lack of or inadequate complaint procedures

      Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.

    4. Lack of or inadequate DHR procedures

      Procedures for device history records have not been established.

    5. DMR - not or inadequately maintained

      A device master record has not been maintained.

    6. Lack of or inadequate final acceptance procedures

      Procedures for finished device acceptance have not been established.

    7. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been established.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Three requests to FDA under the Freedom of Information Act named this site, most recently on Apr 14, 2026. Each one asked for a Form 483.

  • Inspection ending Mar 6, 2024VAI

    2 requests. In the inspection history above. None of FDA’s log entries for these requests say records were released.

    1. Oct 30, 2024FDA-FOIA-2024-9628Form 483· 1 of 3 sitesClosed
    2. Jun 27, 2024FDA-FOIA-2024-5642Form 483Closed
  • Inspection ending Feb 13, 2020NAI

    1 request. In the inspection history above. FDA released records in answer to a request for its Form 483 on Aug 27, 2026. FDAgent holds no copy that FDA posted.

    1. Apr 14, 2026FDA-FOIA-2026-3769Form 483Records released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.

Regulations FDA cited here

The 13 sections of 21 CFR in FDA’s published citations for this facility’s 6 inspections, with how many of them cited each.

Show all 13

Warning letters and compliance actions

FDA Data Dashboard compliance actions

  • Warning letterFormal notice of significant violationsCDRH · DevicesFDA record

FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3010089756 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3010089756. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.