Gunnell Llc
321 W 12th St, Elk City, Oklahoma, United States
FEI 3009255757
FDA has 4 inspections on file for Gunnell Llc in Elk City, Oklahoma (FEI 3009255757), from 2010 to 2016. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below.
The most recent OAI inspection ended on Jun 30, 2010. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Five regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 4, 2010–2016
- 1 OAI · 3 VAI · 0 NAI
- Inspections with 483 observations shown
- 4 of 4
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Lack of or inadequate process validation
A process whose results cannot be fully verified by subsequent inspection and test has not been adequately validated according to established procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.
Report of risk to health
A correction or removal, conducted to reduce a risk to health posed by a device, was not reported in writing to FDA.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Show 1 more citation
Maintenance schedule, Lack of or inadequate schedule
Schedules for the adjustment, cleaning, and other maintenance of equipment have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals to determine whether the quality system activities and results comply with quality system procedures.
Training records
Personnel training is not documented.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Show 6 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated, inspected, checked, and maintained have not been established.
Lack of or inadequate DHR procedures
Procedures for device history records have not been established.
Procedures not adequately established or maintained
Document control procedures have not been adequately established.
Regulations FDA cited here
The 14 sections of 21 CFR in FDA’s published citations for this facility’s 4 inspections, with how many of them cited each.
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 806.10Reports of corrections and removals2 of 4
- 21 CFR 820.50Purchasing controls2 of 4
- 21 CFR 820.90Nonconforming product2 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?1 of 4
- 21 CFR 820.20Management responsibility1 of 4
- 21 CFR 820.22Quality audit1 of 4
Show all 14
- 21 CFR 820.25Personnel1 of 4
- 21 CFR 820.40[Reserved]1 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
- 21 CFR 820.75Process validation1 of 4
- 21 CFR 820.184Device history record1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3009255757 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3009255757. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.