Becton Dickinson and Company
1022 Corporate Park Dr, Mebane, North Carolina, United States
FEI 3008007472
FDA has 5 inspections on file for Becton Dickinson and Company in Mebane, North Carolina (FEI 3008007472), from 2011 to 2023. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
No warning letter, court action or import refusal is filed under this FEI.
Four regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 5, 2011–2023
- 0 OAI · 3 VAI · 2 NAI
- Inspections with 483 observations shown
- 3 of 5
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been documented.
Design verification - documentation
The design verification results, including, were not adequately documented in the design history file.
Nonconforming product control
Products that do not conform to specifications are not adequately controlled.
Equipment control activity documentation
Equipment checks and maintenance activities have not been documented.
Design validation - documentation
The results of design validation, including identification of the design and method(s), were not adequately documented in the design history file.
Show 3 more citations
Quality Audits - defined intervals
Quality audits were not performed at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.
Lack of or inadequate procedures for storage.
Procedures for the control of storage areas and stock rooms have not been established.
Process changes - review, evaluation and revalidation
A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.
- NAINo objectionable conditions that need follow-upDevices · Compliance: DevicesNone
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices6 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Design Validation - Risk analysis not performed/inadequate
Risk analysis is inadequate and is incomplete.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated and maintained have not been adequately established.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
Process changes - review, evaluation and revalidation
A validated process was not reviewed and evaluated and revalidated when changes or process deviations occurred.
Show 1 more citation
Investigation of device failures
Complaints involving the possible failure of a device to meet any of its specifications were not reviewed, evaluated, and investigated where necessary.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Postmarket Assurance: Devices3 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Justification for not reporting
A justification for not reporting the correction or removal action to FDA that included conclusions, follow-ups, and reviews by a designated person was not included in the record.
Time to report - 10 days
A report of the required information regarding device correction and removal actions was not sent to FDA within 10 days of initiating the correction or removal.
Records not kept
There is no record maintained of a correction or removal action that was not required to be reported to FDA.
Regulations FDA cited here
The 11 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.72Inspection, measuring, and test equipment2 of 3
- 21 CFR 820.75Process validation2 of 3
- 21 CFR 806.10Reports of corrections and removals1 of 3
- 21 CFR 806.20Records of corrections and removals not required to be reported1 of 3
- 21 CFR 820.20Management responsibility1 of 3
- 21 CFR 820.22Quality audit1 of 3
- 21 CFR 820.90Nonconforming product1 of 3
Show all 11
- 21 CFR 820.100Corrective and preventive action1 of 3
- 21 CFR 820.150Storage1 of 3
- 21 CFR 820.198Complaint files1 of 3
Other FEIs recorded as Becton Dickinson and Company
Matched on the exact firm name FDA recorded. Sites recorded under other names are not listed, and a shared name does not show common ownership.
Becton Dickinson and CompanyFEI 3007886372San Diego, California, United States
Last inspected 2023-01-11NAI
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3008007472 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3008007472. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.