Skip to content

Pensar Medical

109 N Post Oak Ln Ste 525, Houston, Texas, United States

FEI 3007793449

FDA has 4 inspections on file for Pensar Medical in Houston, Texas (FEI 3007793449), from 2012 to 2026. The latest ended on and was classified OAI, official action indicated. FDA issued a warning letter on (CDRH).

OAI means FDA found significant objectionable conditions and recommended regulatory or administrative action. It is the most serious of FDA's three inspection outcomes.

FDA published 7 citations from this inspection. They are listed below. Two FOIA requests have asked FDA for this inspection's 483.

No court action or import refusal is filed under this FEI.

Two regulations were cited at more than one inspection: 21 CFR 820.100(b) and 21 CFR 820.90(a).

Sources:FDA firm profile ↗Warning letter ↗Synced Nov 23, 2026

Ask FDAgent about this site

Free account: 3 questions, no card needed. Each answer links to its FDA sources.

Email me when FDA posts a new inspection, Form 483 or warning letter for this site

Weekly, on a free account. Pause it any time.

Watch this facility

Comparing suppliers? Check this site beside 2 others, free

Latest inspection
OAI
Inspections on file
4, 2012–2026
1 OAI · 2 VAI · 1 NAI
Inspections with 483 observations shown
3 of 4
Form 483s posted by FDA
None held for this FEI
Requested under FOIA
2 requests, latest Sep 15, 2026
Import refusals
None under this FEI
Import alerts
Not linked by FEI
FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗

Inspection history

4 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC

  1. OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. ISO 13485:2016 Clause 8.2.2

      Complaint handling procedures lack required elements

      Procedures for timely complaint handling do not adequately include requirements and responsibilities for investigating complaints, determining the need to report the information to FDA, handling of complaint-related product and determining the need to initiate corrections or corrective actions.

    2. Lack of System for Event Evaluations

      The written MDR Procedure does not include an internal system which provides for timely and effective identification, communication and evaluation of events that may be subject to medical device reporting requirements.

    3. ISO 13485:2016 Clause 7.3.9

      Requirements for design and development changes not met before implementation

      Design and development changes were not adequately reviewed, verified, validated, and approved before implementation.

    4. ISO 13485:2016 Clause 4.1.2

      Risk-based approach not applied to QMS processes

      A risk-based approach to the control of the appropriate processes needed for the quality management system has not been adequately applied.

    5. ISO 13485:2016 Clause 7.3.6

      Records of design and development verification not maintained

      Records of the results and conclusions of design and development verification and necessary actions were not adequately maintained.

    Show 2 more citations
    1. ISO 13485:2016 Clause 7.3.4

      Records of design and development outputs not maintained

      Records of design and development outputs were not adequately maintained.

    2. ISO 13485:2016 Clause 8.5.2

      Corrective action procedures not documented

      Corrective action procedures defining the requirements for determining the causes of nonconformities, evaluating the need for action to ensure that nonconformities do not recur, planning and documenting action needed and implementing such action, verifying that the corrective action does not adversely affect the ability to meet applicable regulatory requirements or the safety and performance of the medical device and reviewing the effectiveness of corrective action taken have not been adequately documented.

  2. NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices6 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Nonconforming product control

      Products that do not conform to specifications are not adequately controlled.

    2. Documentation

      Corrective and preventive action activities and/or results have not been adequately documented.

    3. Design history file

      The design history file was not established.

    4. Process changes - review, evaluation and revalidation

      A validated process was not revalidated when changes or process deviations occurred.

    5. Acceptable supplier records, inadequate records

      Records of acceptable consultants have not been established.

    Show 1 more citation
    1. Equipment control activity documentation

      Equipment inspections, checks and maintenance activities have not been documented.

  3. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
  4. VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices7 citations published

    FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.

    1. Design Validation - Risk analysis not performed/inadequate

      Risk analysis is inadequate.

    2. Individual Report of Malfunction

      An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.

    3. Investigation of device failures

      Complaints involving the possible failure of a device to meet any of its specifications were not investigated where necessary.

    4. Nonconforming product, Lack of or inadequate procedures

      Procedures have not been adequately established to control product that does not conform to specified requirements.

    5. Documentation

      Corrective and preventive action activities and/or results have not been documented.

    Show 2 more citations
    1. Lack of or inadequate procedures

      Procedures for corrective and preventive action have not been adequately established.

    2. Documentation

      In-process inspections, tests, or other verification activities and approvals were not documented.

Requested under FOIA

FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC

Two requests to FDA under the Freedom of Information Act named this site, most recently on Sep 15, 2026. Each one asked for a Form 483.

  • Inspection ending Apr 28, 2026OAI

    2 requests. In the inspection history above. FDA released records in answer to a request for its Form 483 on Sep 9, 2026. FDAgent holds no copy that FDA posted.

    1. Sep 15, 2026FDA-FOIA-2026-8918Form 483· 1 of 5 sitesRecords released
    2. Jun 19, 2026FDA-FOIA-2026-5953Form 483· 1 of 5 sitesRecords released

A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out.

Regulations FDA cited here

The 10 sections of 21 CFR in FDA’s published citations for 3 of this facility’s 4 inspections, with how many of them cited each.

Show all 10

Warning letters and compliance actions

FDA Data Dashboard compliance actions

Go further than this page

FDAgent researches FDA’s records and links every answer to its source.

Free account: 3 questions, no card needed.

About this data. Every record here is filed under FEI 3007793449 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
  • Every record on this page is filed under FEI 3007793449. Records filed under another FEI or name, including other sites of the same company, are not shown here.
  • FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
  • Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
  • “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
  • Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
  • FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.