Glasir Medical, LP
9862 Lorene Ln Ste 108, San Antonio, Texas, United States
FEI 3007789227
FDA has 3 inspections on file for Glasir Medical, LP in San Antonio, Texas (FEI 3007789227), from 2012 to 2017. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 13 citations from this inspection: 11 under 21 CFR Part 820, the device quality system rules, 1 under 21 CFR Part 803, the device adverse event reporting rules, and 1 under 21 CFR Part 1271, the rules for human cell and tissue products. They are listed below.
No warning letter, court action or import refusal is filed under this FEI.
Two regulations were cited at more than one inspection: 21 CFR 820.20(b) and 21 CFR 820.22.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 3, 2012–2017
- 0 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 3 of 3
- Form 483s posted by FDA
- None held for this FEI
- Warning letters
- None on file
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
3 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies13 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been adequately established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been adequately established.
Design input - Lack of or inadequate procedures
Procedures for design input have not been established.
Design validation - user needs and intended uses
Design validation did not ensure the device conforms to defined user needs and intended uses.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed in accordance with 21 CFR 820.
Show 8 more citations
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been adequately established.
Individual Report of Malfunction
An MDR report was not submitted within 30 days of receiving or otherwise becoming aware of information that reasonably suggests that a marketed device has malfunctioned and would be likely to cause or contribute to a death or serious injury if the malfunction were to recur.
Lack/Inad procedure-Monitoring/Control of Validated Proces
Procedures for monitoring and control of process parameters for a validated process have not been established.
Procedures and release criteria
Procedures including release criteria for activities relating to the receipt of HCT/Ps were not implemented.
Design validation - Risk analysis
Results of the design risk analysis were not adequately documented.
Design changes - Lack of or Inadequate Procedures
Procedures for design change have not been adequately established.
Quality Audits - defined intervals
Quality audits were not performed at defined intervals and at sufficient frequency to determine whether the quality system activities and results comply with quality system procedures.
Design review - Lack of or inadequate procedures
Procedures for design review have not been adequately established.
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Design verification - output does not meet input requirement
Design verification does not confirm that design output meets design input requirements.
Supplier oversight
The type and extent of control to be exercised over suppliers, contractors, and consultants was not clearly defined.
Quality audits - Lack of or inadequate procedures
Procedures for quality audits have not been adequately established.
Documentation
Process validation activities and results have not been documented and approved.
DMR - not or inadequately maintained
A device master record has not been adequately maintained.
Show 3 more citations
Management review - Lack of or inadequate procedures
Procedures for management review have not been adequately established.
Incorrect translation to production specifications
The device design was not correctly translated into production specifications.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are produced in accordance with 21 CFR 820.
- VAIObjectionable conditions found; FDA took no regulatory actionBiologics · Human Cellular, Tissue, and Gene Therapies4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures to meet core CTGP
Procedures appropriate to meet core CGTP requirements for all steps that you perform in the manufacture of HCT/Ps were not established.
Storage temperatures recorded, maintained
Storage temperatures of HCT/Ps were not recorded.
Return to inventory--procedures
Procedures for determining if HCT/Ps that were returned to the establishment are suitable to be returned to inventory were not established.
Procedures re complaints
Procedures for the review, evaluation, documentation, and investigation of complaints relating to core CGTP requirements were not established.
Regulations FDA cited here
The 13 sections of 21 CFR in FDA’s published citations for this facility’s 3 inspections, with how many of them cited each.
- 21 CFR 820.20Management responsibility2 of 3
- 21 CFR 820.22Quality audit2 of 3
- 21 CFR 820.30Design controls2 of 3
- 21 CFR 820.50Purchasing controls2 of 3
- 21 CFR 820.75Process validation2 of 3
- 21 CFR 1271.265Receipt, predistribution shipment, and distribution of an HCT/P2 of 3
- 21 CFR 803.50If I am a manufacturer, what reporting requirements apply to me?1 of 3
- 21 CFR 820.100Corrective and preventive action1 of 3
Show all 13
- 21 CFR 820.181Device master record1 of 3
- 21 CFR 820.198Complaint files1 of 3
- 21 CFR 1271.180Procedures1 of 3
- 21 CFR 1271.260Storage1 of 3
- 21 CFR 1271.320Complaint file1 of 3
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3007789227 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3007789227. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.