21 CFR 1271.265: Receipt, predistribution shipment, and distribution of an HCT/P
FDA's inspection data lists 261 Form 483 citations of 21 CFR 1271.265 at 176 facilities, from October 2008 to August 27, 2025.
Counts: FDA Data Dashboard citations, last synced Nov 23, 2026, 04:25 UTC, and FDA's warning letters, last synced Nov 23, 2026, 04:17 UTC. Regulation text: eCFR. FDAgent is independent of FDA.
- Form 483 citations
- 261
- At 194 inspections
- Facilities cited
- 176
- Distinct FEIs
- Most recent citation
- Aug 27, 2025
- Lightbody Medical Technologies Inc dba Artemis Surgical
- Warning letters
- 0
- State a violation of it
The regulation
The current text of 21 CFR 1271.265 from the Electronic Code of Federal Regulations.
§ 1271.265 Receipt, predistribution shipment, and distribution of an HCT/P.
(a) Receipt. You must evaluate each incoming HCT/P for the presence and significance of microorganisms and inspect for damage and contamination. You must determine whether to accept, reject, or place in quarantine each incoming HCT/P, based upon pre-established criteria designed to prevent communicable disease transmission.
(b) Predistribution shipment. If you ship an HCT/P within your establishment or between establishments (e.g., procurer to processor) and the HCT/P is not available for distribution as described in paragraph (c) of this section, you must first determine and document whether pre-established criteria designed to prevent communicable disease transmission have been met, and you must ship the HCT/P in quarantine.
(c) Availability for distribution. (1) Before making an HCT/P available for distribution, you must review manufacturing and tracking records pertaining to the HCT/P, and, on the basis of that record review, you must verify and document that the release criteria have been met. A responsible person must document and date the determination that an HCT/P is available for distribution.
(2) You must not make available for distribution an HCT/P that is in quarantine, is contaminated, is recovered from a donor who has been determined to be ineligible or for whom a donor-eligibility determination has not been completed (except as provided under §§ 1271.60, 1271.65, and 1271.90), or that otherwise does not meet release criteria designed to prevent communicable disease transmission.
(3) You must not make available for distribution any HCT/P manufactured under a departure from a procedure relevant to preventing risks of communicable disease transmission, unless a responsible person has determined that the departure does not increase the risk of communicable disease through the use of the HCT/P. You must record and justify any departure from a procedure at the time of its occurrence.
(d) Packaging and shipping. Packaging and shipping containers must be designed and constructed to protect the HCT/P from contamination. For each type of HCT/P, you must establish appropriate shipping conditions to be maintained during transit.
(e) Procedures. You must establish and maintain procedures, including release criteria, for the activities in paragraphs (a) through (d) of this section. You must document these activities. Documentation must include:
(1) Identification of the HCT/P and the establishment that supplied the HCT/P;
(2) Activities performed and the results of each activity;
(3) Date(s) of activity;
(4) Quantity of HCT/P subject to the activity; and
(5) Disposition of the HCT/P (e.g., identity of consignee).
(f) Return to inventory. You must establish and maintain procedures to determine if an HCT/P that is returned to your establishment is suitable to be returned to inventory.
21 CFR Part 1271: Human Cells, Tissues, and Cellular and Tissue-Based Products, Subpart DRead on eCFR
Citations by year
Form 483 citations of 21 CFR 1271.265 by the year the inspection ended. FDA's citation data starts with inspections ending in October 2008, and the current year runs to the latest sync, so both are partial years (lighter bars).
Peak: 37 in 2012
| Year | Form 483 citations |
|---|---|
| 2008 | 8 |
| 2009 | 28 |
| 2010 | 18 |
| 2011 | 21 |
| 2012 | 37 |
| 2013 | 16 |
| 2014 | 32 |
| 2015 | 14 |
| 2016 | 22 |
| 2017 | 9 |
| 2018 | 6 |
| 2019 | 11 |
| 2021 | 3 |
| 2022 | 10 |
| 2023 | 4 |
| 2024 | 10 |
| 2025 | 12 |
How FDA words this citation
FDA's short descriptions and citation text on these 261 citations, most used first. These are FDA's standard summaries in its citation data, not the text of any one Form 483.
Short descriptions
- Procedures and release criteria75 citations
- Return to inventory--procedures68 citations
- Shipping conditions appropriate29 citations
- Documentation elements for activities26 citations
- Evaluation--microorganisms, damage19 citations
Show 3 more
- Acceptance criteria designed to prevent CD15 citations
- Improperly released for distribution8 citations
- Release criteria verified, documented8 citations
Citation text
- “Procedures for determining if HCT/Ps that were returned to the establishment are suitable to be returned to inventory were not established.”20 citations
- “Procedures including release criteria for activities relating to the receipt, shipment and distribution of HCT/Ps were not established, maintained, documented and implemented.”11 citations
- “Incoming HCT/Ps were not evaluated for the presence and significance of microorganisms and not inspected for damage and contamination.”10 citations
- “Incoming HCT/Ps were not accepted, rejected and placed in quarantine based on pre-established criteria designed to prevent communicable disease transmission.”9 citations
Show 2 more
- “Procedures for determining if HCT/Ps that were returned to the establishment are suitable to be returned to inventory were not established, maintained, defined and documented.”8 citations
- “Procedures including release criteria for activities relating to the receipt, shipment and distribution of HCT/Ps were not established.”8 citations
Facilities cited most recently
The latest inspections where FDA cited 21 CFR 1271.265, by inspection end date, with FDA's short description. Each links to the facility's FDA record on FDAgent.
- Lightbody Medical Technologies Inc dba Artemis Surgical2025-08-27 · FEI 3007048062Evaluation--microorganisms, damage
Show 4 more
Often cited at the same inspection
Of the 194 inspections where FDA cited 21 CFR 1271.265, the share where it also cited each of these sections.
- 21 CFR 1271.260Storage82 (42%)
- 21 CFR 1271.180Procedures61 (31%)
- 21 CFR 1271.320Complaint file51 (26%)
- 21 CFR 1271.160Establishment and maintenance of a quality program49 (25%)
- 21 CFR 1271.270Records32 (16%)
- 21 CFR 1271.200Equipment26 (13%)
Ask FDAgent about 21 CFR 1271.265
FDAgent reads FDA's Form 483 citations, observations and warning letters and answers in plain English, with links to the records.
Related
About this data
- Citation counts come from FDA's Data Dashboard, which lists the regulation behind each Form 483 observation for inspections since October 2008.
- Observation excerpts come from Form 483 PDFs in FDA's electronic reading room, tied to this section only by exact wording as described above.
- A warning letter counts for this section when one of its numbered violation sentences cites 21 CFR 1271.265; FDAgent quotes that sentence exactly.
- FDAgent rebuilds this page every week from FDA's public data and is independent of FDA.