Asept Pak Inc.
64 West St, Malone, New York, United States
FEI 3007177643
FDA has 6 inspections on file for Asept Pak Inc. in Malone, New York (FEI 3007177643), from 2011 to 2023. The latest ended on and was classified NAI, no action indicated.
NAI means FDA found no objectionable conditions or practices, or found none that justified further action.
FDA published 2 citations from this inspection, all under 21 CFR Part 820, the device quality system rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
Seven regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- NAI
- Inspections on file
- 6, 2011–2023
- 0 OAI · 4 VAI · 2 NAI
- Inspections with 483 observations shown
- 5 of 6
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Oct 23, 2023
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
6 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices2 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Periodic equipment inspections
Periodic inspections of equipment were not conducted to ensure adherence to applicable maintenance schedules.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance7 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Procedures not in writing, fully followed
The responsibilities and procedures applicable to the quality control unit are not in writing and fully followed.
Cleaning / Sanitizing / Maintenance
Equipment and utensils are not maintained at appropriate intervals to prevent malfunctions and contamination that would alter the safety, identity, strength, quality or purity of the drug product.
Expiration date lacking
Drug products do not bear an expiration date determined by appropriate stability data to assure they meet applicable standards of identity, strength, quality and purity at the time of use.
Control procedures to monitor and validate performance
Control procedures are not established which monitor the output and validate the performance of those manufacturing processes that may be responsible for causing variability in the characteristics of in-process material and the drug product.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established.
Show 2 more citations
Plumbing System Defects
The plumbing system contains defects that could contribute to the contamination of drug products.
Records reviewed annually
Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Expiration date lacking
Drug products do not bear an expiration date determined by appropriate stability data to assure they meet applicable standards of identity, strength, quality and purity at the time of use.
Contamination control, Lack of or inadequate procedures
Procedures to prevent contamination of equipment or product by substances that may have an adverse effect on product quality have not been adequately established.
Environmental control Lack of or inadequate procedures
Procedures to control environmental conditions have not been adequately established.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: DevicesCitations not published
- NAINo objectionable conditions that need follow-upDevices · Postmarket Assurance: Devices11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
SOPs not followed / documented
Written production and process control procedures are not followed in the execution of production and process control functions and documented at the time of performance.
Design validation - documentation
The results of design validation, including identification of the design, method(s), the date, and the individual(s) performing validation, were not documented in the design history file.
Expiration date lacking
Drug products do not bear an expiration date determined by appropriate stability data to assure they meet applicable standards of identity, strength, quality and purity at the time of use.
In-process materials specifications testing
Examination and testing of samples is not done to assure that in-process materials conform to specifications.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established and written.
Show 6 more citations
Equipment for Environmental Control
Equipment for adequate control over humidity and temperature is not provided when appropriate for the manufacture, processing, packing or holding of a drug product.
Lack of written stability program
There is no written testing program designed to assess the stability characteristics of drug products.
Documentation
Process validation activities and results have not been documented.
Design history file
The design history file does not demonstrate that the design was developed following the requirements of 21 CFR 820.
Documentation
Corrective and preventive action activities and/or results have not been adequately documented.
Records reviewed annually
Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures.
- VAIObjectionable conditions found; FDA took no regulatory actionDrugs · Drug Quality Assurance11 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Defined areas of adequate size for operations
The separate or defined areas necessary to prevent contamination or mix-ups are deficient.
Retention time of reserve samples, in general
Reserve samples for active ingredients are not retained for one year after the expiration date of the drug product.
Absence of Written Procedures
There are no written procedures for production and process controls designed to assure that the drug products have the identity, strength, quality, and purity they purport or are represented to possess.
Lack of written stability program
There is no written testing program designed to assess the stability characteristics of drug products.
In-process materials specifications testing
Examination and testing of samples is not done to assure that in-process materials conform to specifications.
Show 6 more citations
Expiration date lacking
Drug products do not bear an expiration date determined by appropriate stability data to assure they meet applicable standards of identity, strength, quality and purity at the time of use.
Acceptance criteria for sampling & testing
Acceptance criteria for the sampling and testing conducted by the quality control unit is not adequate to assure that batches of drug products meet appropriate statistical quality control criteria as a condition for their approval and release.
Records reviewed annually
Records are not maintained so that data therein can be reviewed at least annually to evaluate the quality standards of each drug product to determine the need for changes in specifications or manufacturing or control procedures.
Procedures for sterile drug products
Procedures designed to prevent microbiological contamination of drug products purporting to be sterile are not established and written.
Equipment for Environmental Control
Equipment for adequate control over humidity and temperature is not provided when appropriate for the manufacture, processing, packing or holding of a drug product.
Buildings not maintained in good state of repair
Buildings used in the manufacturing of a drug product are not maintained in a good state of repair.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Oct 23, 2023. It asked for a Form 483.
Inspection ending Aug 4, 2023NAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Oct 23, 2023FDA-FOIA-2023-9328Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 18 sections of 21 CFR in FDA’s published citations for 5 of this facility’s 6 inspections, with how many of them cited each.
- 21 CFR 211.137Expiration dating4 of 5
- 21 CFR 211.110Sampling and testing of in-process materials and drug products3 of 5
- 21 CFR 211.113Control of microbiological contamination3 of 5
- 21 CFR 211.180General requirements3 of 5
- 21 CFR 820.100Corrective and preventive action3 of 5
- 21 CFR 211.46Ventilation, air filtration, air heating and cooling2 of 5
- 21 CFR 211.100Written procedures; deviations2 of 5
- 21 CFR 211.166Stability testing2 of 5
Show all 18
- 21 CFR 820.70Production and process controls2 of 5
- 21 CFR 211.22Responsibilities of quality control unit1 of 5
- 21 CFR 211.42Design and construction features1 of 5
- 21 CFR 211.48Plumbing1 of 5
- 21 CFR 211.58Maintenance1 of 5
- 21 CFR 211.67Equipment cleaning and maintenance1 of 5
- 21 CFR 211.165Testing and release for distribution1 of 5
- 21 CFR 211.170Reserve samples1 of 5
- 21 CFR 820.30Design controls1 of 5
- 21 CFR 820.75Process validation1 of 5
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3007177643 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3007177643. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.