Apolo - Megasun LLC
4515 Miami St, Saint Louis, Missouri, United States
FEI 3006190476
FDA has 5 inspections on file for Apolo - Megasun LLC in Saint Louis, Missouri (FEI 3006190476), from 2010 to 2024. The latest ended on and was classified VAI, voluntary action indicated.
VAI means FDA investigators found objectionable conditions or practices, but FDA was not prepared to take or recommend regulatory action. The firm is expected to correct them voluntarily.
FDA published 4 citations from this inspection: 3 under 21 CFR Part 820, the device quality system rules, and 1 under 21 CFR Part 801, the device labeling rules. They are listed below. One FOIA request has asked FDA for this inspection's 483.
The most recent OAI inspection ended on Mar 6, 2013. FDA issued a warning letter on (CDRH). No court action or import refusal is filed under this FEI.
12 regulations were cited at more than one inspection; they are listed below.
Sources:FDA firm profile ↗Synced Nov 23, 2026
Ask FDAgent about this site
Email me when FDA posts a new inspection, Form 483 or warning letter for this site
Weekly, on a free account. Pause it any time.
Comparing suppliers? Check this site beside 2 others, free
- Latest inspection
- VAI
- Inspections on file
- 5, 2010–2024
- 2 OAI · 2 VAI · 1 NAI
- Inspections with 483 observations shown
- 4 of 5
- Form 483s posted by FDA
- None held for this FEI
- Requested under FOIA
- 1 request, latest Nov 1, 2024
- Warning letters
- · CDRH ↗
- Import refusals
- None under this FEI
- Import alerts
- Not linked by FEI
- FDA lists import alert firms by name and address, rarely by FEI. Check FDA’s import alert list ↗
Inspection history
5 inspections · FDA Data Dashboard, last synced Nov 23, 2026, 04:25 UTC
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices4 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Calibration, Inspection, etc. Procedures Lack of or Inadequ
Procedures to ensure equipment is routinely calibrated have not been adequately established.
Label to bear a unique device identifier
The labels of the medical device(s) do not bear a unique device identifier (UDI) that meet the requirements of 801 Subpart B and 21 CFR 830.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
Design control - no procedures
Procedures for design control have not been established.
- NAINo objectionable conditions that need follow-upDevices · Radiation Control and Health Safety ActNone
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Compliance: Devices8 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
DHR - not or inadequately maintained
A device history record has not been adequately maintained.
Show 3 more citations
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been adequately established.
- VAIObjectionable conditions found; FDA took no regulatory actionDevices · Compliance: Devices16 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Purchasing controls, Lack of or inadequate procedures
Procedures to ensure that all purchased or otherwise received product and services conform to specified requirements have not been established.
Quality plan
A quality plan has not been established.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Lack of or inadequate final acceptance procedures
Procedures for finished device acceptance have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Show 11 more citations
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
Failure to submit, distinct marking
You did not submit a required product report prior to the introduction of the product into commerce.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Lack of or inadequate organizational structure
The organizational structure has not been adequately established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Training - Lack of or inadequate procedures
Procedures for training and identifying training needs have not been established.
Management review dates
The results and/or dates of management reviews are not documented.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Process control procedures, Lack of or inadequate procedures
Process control procedures that describe any process controls necessary to ensure conformance to specifications have not been established.
Lack of or inadequate procedures - Acceptance activities
Procedures for acceptance activities have not been established.
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
- OAISignificant objectionable conditions; FDA recommended regulatory actionDevices · Postmarket Assurance: Devices12 citations published
FDA’s citation titles and text from its Data Dashboard. These are FDA’s summaries, not the Form 483 itself.
Management ensuring quality policy is understood
Management with executive responsibility has not ensured that the quality policy is understood, implemented and maintained at all levels of the organization.
Nonconforming product, Lack of or inadequate procedures
Procedures have not been established to control product that does not conform to specified requirements.
Failure to submit, distinct marking
You did not submit a required product report prior to the introduction of the product into commerce.
Management review - Lack of or inadequate procedures
Procedures for management review have not been established.
Lack of Written MDR Procedures
Written MDR procedures have not been developed.
Show 7 more citations
Quality plan
A quality plan has not been established.
Lack of or inadequate procedures
Procedures for corrective and preventive action have not been established.
Quality Audit/Reaudit - conducted
Quality audits have not been performed.
Management review dates
The results and/or dates of management reviews are not documented.
Failure to submit
You did not submit an annual report for products requiring one.
Lack of or inadequate organizational structure
The organizational structure has not been established and maintained to ensure that devices are designed and produced in accordance with 21 CFR 820.
Lack of or inadequate complaint procedures
Procedures for receiving, reviewing, and evaluating complaints by a formally designated unit have not been established.
Requested under FOIA
FDA FOIA logs, last synced Nov 23, 2026, 04:25 UTC
One request to FDA under the Freedom of Information Act named this site, on Nov 1, 2024. It asked for a Form 483.
Inspection ending Aug 29, 2024VAI
1 request. In the inspection history above. None of FDA’s log entries for these requests say records were released.
- Nov 1, 2024FDA-FOIA-2024-9659Form 483Closed
A request is listed when it names this site’s FEI number or one of its FDA inspection IDs, or, after review, its exact street address. Requester names are left out. “Closed” means FDA’s log gives no further reason.
Regulations FDA cited here
The 16 sections of 21 CFR in FDA’s published citations for 4 of this facility’s 5 inspections, with how many of them cited each.
- 21 CFR 820.20Management responsibility3 of 4
- 21 CFR 820.22Quality audit3 of 4
- 21 CFR 820.80Receiving, in-process, and finished device acceptance3 of 4
- 21 CFR 820.100Corrective and preventive action3 of 4
- 21 CFR 820.198Complaint files3 of 4
- 21 CFR 803.17What are the requirements for developing, maintaining, and implementing written MDR procedures that apply to me?2 of 4
- 21 CFR 820.25Personnel2 of 4
- 21 CFR 820.90Nonconforming product2 of 4
Show all 16
- 21 CFR 1002.10Product reports2 of 4
- 21 CFR 801.20Label to bear a unique device identifier1 of 4
- 21 CFR 820.30Design controls1 of 4
- 21 CFR 820.50Purchasing controls1 of 4
- 21 CFR 820.70Production and process controls1 of 4
- 21 CFR 820.72Inspection, measuring, and test equipment1 of 4
- 21 CFR 820.184Device history record1 of 4
- 21 CFR 1002.13Annual reports1 of 4
Warning letters and compliance actions
FDA Data Dashboard compliance actions
- Warning letterFormal notice of significant violationsCDRH · DevicesFDA record
FDAgent links a letter’s text only when FDA’s compliance case number ties it to this FEI. No posted letter matched this one, so the link opens FDA’s record of the action.
Go further than this page
FDAgent researches FDA’s records and links every answer to its source.
About this data. Every record here is filed under FEI 3006190476 in FDA’s data. FDAgent is independent of FDA, and FDA’s firm profile is the authoritative record.
- Every record on this page is filed under FEI 3006190476. Records filed under another FEI or name, including other sites of the same company, are not shown here.
- FDA posts only some Form 483s. A missing 483 does not mean no observations were issued, and classifications can change after FDA review.
- Recalls are not shown, because FDAgent does not hold them by FEI. Import alerts are shown when FDA printed this FEI beside the firm on the alert, or when a Red List entry’s firm name, street address and country equal an inspection record under this FEI and no other FEI has that name. FDA lists most firms by name and address alone, so a site can be on an import alert without it appearing here.
- “Last synced” is when FDAgent last copied that FDA source. FDA often publishes a record weeks after the inspection or action it describes.
- Observation text comes from OCR of the scanned PDF. FDAgent keeps an observation only when it matches the scanned text, and marks unreadable spans [illegible]. Last synced Nov 23, 2026, 04:30 UTC.
- FDAgent is independent and not affiliated with FDA. Check the FDA firm profile for the authoritative record.